VIII– USING THE DEVICE
Before turning on the central unit, check to see that you have completed the installation of
ALL COMPONENTS.
Important:
- Do not disconnect the micro-motor cord when the device is switched on and the footswitch is pressed.
- Do not engage the instrument to or disengage it from the contra-angle when the micromotor is running.
- All accessories must be cleaned, disinfected and sterilized before use.
- For your safety and that of your patient, the I-SURGE must not be used with accessories other than those
supplied by BY DENTAL and/or SATELEC.
- Check the integrity of the device and its accessories before and after each use in order to detect any problem.
- If necessary, do not use the device and replace any faulty component.
The I-SURGE is prepared for use as follows:
- Make sure that the device is connected correctly and properly insulated.
- Micromotor connection:
Connect the micro motor into the receptacle on the front of the central unit (Right – white – 8
Pins) (Fig. 2 -10), (Fig. 8).
Connect the "E-type" ISO 3964 contra-angle to the micro motor and fit an instrument (drill, bur,
etc.) to the contra-angle.
Place the micro motor on the motor holder.
- Footswitch connection:
Connect the footswitch into the receptacle on the front of the central unit (Left – Blue –5 Pins)
(Fig. 2 – 11).
Place the footswitch on the ground.
While inserting the plug, align the guide on the connector in an upright position to fit into the
corresponding female guide on the I-SURGE. Connectors have a spring retention system to help
avoid accidentally disconnecting the unit's cords and cables.
- Irrigation connection:
- Hang the bottle or bag of physiological saline from the bracket.
- Insert the irrigation line to the pump ("IN" = perforator side ; "OUT" = contra-angle side) and insert the
irrigation line perforating pin into the bottle or bag of physiological saline (Fig. 10).
- Fasten the irrigation line to the cord using line clips.
- Connect the end of the irrigation line to the contra-angle.
Note: This sequence must be completed in a controlled asepsis context for the safety of your patients. An
operating assistant may be needed.
The manufacturer accepts no responsibility in the case of irrigation tubes other than those mentioned in this
manual, being used.
- Switch on the device:
- Switch the power switch to position I (power on).
- Make any necessary setting adjustments (irrigation flow rate, speed, torque, program, etc.)
- Use I-SURGE in accordance with good dental practices.
15