Technical Data - ProMinent DULCOTEST CTE 1-mA Mode D'emploi

Cellule de mesure de chlore total
Masquer les pouces Voir aussi pour DULCOTEST CTE 1-mA:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Technical Data

14 Technical Data
Measured
variable
Total chlorine:
Applications
Potable, industrial, process and cooling water,
brine and sea water for swimming pool and Jacuzzi treatment
Measurement
CTE 1-mA-0.5 ppm
range and
CTE 1-mA-2 ppm
standard slope
CTE 1-mA-5 ppm
CTE 1-mA-10 ppm
CTE 1-DMT-10 ppm
Solution
Corresponds to lower measurement range limit
Reaction time
T
pH range
6.5 – 9.5
pH-
dependency
Jump from pH 7 to pH 8: approx. -10 % for free chlorine
Conductivity
water sample
0.03 - 40 mS/cm
Temperature
range
5 - 45 °C, temperature compensation integrated, no quick changes in
temperature allowed
Pressure
sample water in DLG III in-line housing: max. 1 bar (free outlet)
sample water in DGM in-line housing:
Flow
sample water through DLG III and DGM in-line
probe housing:
Cross sensitivity
Oxidising reagents e.g. bromine, iodine, ozone, chlorine dioxide,
permanganates result in false positive results.
Reducing reagents like sulphides, sulphites, thiosulphates, and hydrazine
result in false negative results.
Service life
membrane cap
Typically 3 - 6 months, depending on water quality
46
free chlorine (Cl
bound chlorine (chloramines)
organic bound chlorine (e.g. cyanuric acid derivates)
approx. 60 s (as concentration rises and falls)
90
(dissolved) HOCl, OCl - )
2
0.01 - 0.5 mg/l
standard slope: 24 mA/ppm
0.02 - 2.0 mg/l
standard slope: 6 mA/ppm
0.05 - 5.0 mg/l
standard slope: 2.4 mA/ppm
0.01 - 1.0 mg/l
standard slope: 1.2 mA/ppm
0.01 - 10 mg/l
standard slope: 32.5 mV/ppm
max. 3 bar (no pressure spikes
allowed)
optimum:
minimum:
maximum: 100 l/h
30 l/h
20 l/h
ProMinent
®

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Dulcotest cte 1-dmt

Table des Matières