Electrolyte You can pour the electrolyte down the drain.
Measuring cell
Measured variables Peracetic acid
Applications Particularly suitable for Cleaning in Place (CIP), rinser, PET bottle
Measuring range PAA 1-mA-200 ppm:
Dispersal In accordance with the lower limit of the instrument measuring
Nominal slope conductance With pH 4, T = 30 °C:
Reaction time T
pH range 1 - 9 (peracetic acid stability range)
Temperature range 5 - 45 °C, temperature compensated, no sudden leaps in
Cross-sensitivity Bromine, bromamine, chloride, ozone.
Lifespan of diaphragm cap Typically 3-6 months, depending on water quality
Materials Diaphragm cap:
Power supply 16 - 24 V DC; min 35 mA at 16 V DC
Output signal 4 - 20 mA
Enclosure rating IP 65
Storage temperature Between 5 and 50 °C
ProMinent
®
13
You can dispose of the diaphragm cap with normal household waste.
IMPORTANT
Observe the regulations, which currently apply in your particular
location (especially with regard to electronic scrap!)
14
cleaning
PAA 1-mA-2000 ppm: 10... 2000 mg/l
range
PAA 1-mA-200 ppm:
PAA 1-mA-2000 ppm:
approx. 3 min
90
temperature (sample water), 5 - 50 °C (air temperature)
Pressure Sample water in DLG III in-line probe housing:
Maximum 1 bar (free flow)
Sample water in DGM in-line probe housing:
Maximum 3 bar (30
Flow Sample water through in-line probe housing DLG III, DGM
optimal:
50 l/h minimum: 20 l/h
maximum: 100 l/h
The probe shows no signs of cross-sensitivity to standard aqueous
solutions of hydrogen peroxide.
Electrode shaft:

Disposal

Technical data

1... 200 mg/l
60 µA/ppm
6 µA/ppm
o
C, no sharp pressure peaks permitted)
PVDF
PVC-C
Disposal / Technical Data