If using an alternative controller please note:
Power source: 16-24 V DC, min. 35 mA with 16 V DC
Max. load:
Electrical
installation
CTE 1-mA
Fig. 4
Electrical
connection of
the sensor
CTE 1-mA
CTE 1-DMT
®
ProMinent
1 W
IMPORTANT
For electrical connection of the sensor to the controller, only
use wires with a diameter of 4 mm (see Order information,
Chapter 15)
Turn the top part of the sensor a quarter of a turn anticlockwise and
remove.
Remove the outer isolation of the cable for about 5 cm from the end so
that the two wires appear.
Loosen the M12 threaded connector and guide the 2-wire cable into the
sensor. Leave an approx. 5 cm length of the 2-core measuring line in the
measuring cell.
Strip the two wires at their ends and connect them to the terminal block
by using the screwdriver provided: 1 = plus, 2 = minus (see figure 4).
Tighten up the M12 connector.
Insert the top part of the sensor right into the housing and turn the top
part clockwise as far as it will go.
Insert the 5-pin plug into the socket.
Assembly