B.A. International Powersurge PS3X Notice D'instruction page 11

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 17
V.2 - Connection to the motor
• Check that the MD is completely dry before connecting it to the motor.
x
• Install the MD on the motor connection until ratcheting. For this, keep
the motor and the MD in the same axle.
: Turn the MD until the retractable catch lodges in the E
d
b
notch of the motor (Fig 3).
or
Apply light traction to the MD to check that it is properly
attached to the motor connection.
Operational test by switching on the motor; start at low speed, then
gradually increase.
If you notice overheating, irregularities, vibrations, abnormal noises
when using the MD, immediately contact your AS.
V.3 - Connection of the irrigation (Fig. 4)
x
Attach the spraying nozzle to the head of the MD.
Connect the extremity of the irrigation pipe to the nozzle entry.
V.4 - Disconnection of motor
Remove the MD by keeping it in the motor axle.
In the event of prolonged non use of the device, do not leave it connected to
D
the motor. Risk of damaging the motor.
V.5 - Connection/disconnection of rotary instrument
r
For all handling of rotary instruments, wear protective gloves. Risk of
cutting and contamination.
x
Only use rotary instruments in good condition. Risk of necrosis.
Insertion of rotary instrument
Contra-angle:
• Apply pressure using your thumb to the centre of the push button
and simultaneously insert the rotary tool up to the stop in the claw
of the MD.
• Release the thumb pressure on the push button (Fig 5)
Handpiece:
• Align the p mark to the front of the  mark by turning the front part
in relation to the back part
• Insert the rotary instrument up to the stop.
• Align the p mark to the front of mark l.
For tool replacement: check the good condition of the rotary tool with
E
light axial traction.
EN
FR
ES
IT
DE
PO
11

Publicité

Table des Matières
loading

Table des Matières