Specifications
Filter
- Viral/bacterial filtration efficiency: Greater than 99.999%.
- Particle filtration efficiency: Minimum of 99.97% of 0.3-μm nominal particle size at 100 L/min
flow.
- Inlet connector: 22-mm ISO connector
- Resistance to flow: Less than 2.5 cmH
- Flow leakage: Less than 0.01 L/min at 140 cmH
- Size: 11.7 cm diameter x 12.7 cm long (4.6 in. diameter x 5 in. long).
- Internal volume: Approximately 400 mL.
Collector vial
- Capacity: Approximately 200 mL usable volume (to maximum fill line).
Tubing
- Length: 17.8 cm (7 in.)
- Internal diameter: 0.48 cm (3/16 in.)
Drain bag
- Capacity: Approximately 1 L.
Ordering information
4-070305-00
4-074647-00
4-070315-00
4-048491-00
4-048493-00
4-048492-00
4-074613-00
4-070311-00
O at 100 L/min when new.
2
Expiratory filter, Re/X800, reusable
Collector vial, reusable
Expiratory filter and collector vial, D/X800, single-patient use
(carton of 12)
Drain bag, single-patient use (package of 25)
Tubing, drain bag, single-patient use (package of 10)
Clamp, reusable (package of 5)
Drain cap
Seal, expiratory filter
O internal pressure.
2
6