Technical Data - Amprobe TELARIS Série Mode D'emploi

Masquer les pouces Voir aussi pour TELARIS Série:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

Technical Data

Display:
Duration of measurement:
Measurement principle:
Measurement voltage:
Current/power consumption: approx. 10mA/ 0,1W max. 500mA briefly, during measurement
Measurement current:
Frequency (f1/ f2):
Measurement method:
RE:
RE:
RH, RS:
IM:
*operating error
Nominal use range:
Correct polarity:
Noise voltage:
aux. earth (RH), probe resis. (RS):
Ratio RH/ RE:
Ratio RS/ RE:
Earth resistance (RE):
Reference range:
Correct polarity:
Noise voltage:
aux. earth (RH),
probe resis. (RS)
Earth resistance (RE):
Auto power off
Power supply
Low Bat at
Battery life
Dimensions (LxWxD)
Weight
3 digit LCD, 1999 digit
3-5s
current-voltage measurement principle in accordance with
EN61557-5
25V/ 50V selection
max. 12mA
127Hz/ 140Hz, selectable
2 pole/ 3 pole/ 4 pole
Range
Tolerance
0.05 - 2.00 Ohm ±(4% + 0.1 Ohm)
2.00 - 1999 Ohm ±(4% + 2 Digit)
0.1 - 50 kOhm
±(10% + 3 Digit)*
0.1 - 13mA
±(10% + 3 Digit)*
for 3/4 pole measurements
0...5V, max. 10% of measurement voltage
max. 50 kOhm
max. 1000/1
max. 10000/1
max. 2 kOhm
for 3/4 pole measurements
0 V
0 Ohm
max. 2 kOhm
after approx. 5 min
6 x 1.5 V Type IEC LR6 (Mignon/Size AA)
6.6 V ±5 % (OFF at 6.1V ±5 %)
approx. 1 year for normal use
235x105x68 mm
approx. 640g (inc. 6 x batteries)
22
±(6% + 0.1 Ohm)*
±(6% + 3 Digit)*

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Telaris earth-test

Table des Matières