Technical Data

Insulation Resistance Measurement
Measurement Range
0,010...1,999 MΩ
2,00...19,99 MΩ
20,0...199,9 MΩ
0,200...1,999 GΩ
Battery Life
Nominal test
Voltage UN
100 V DC
250 V DC
500 V DC
1000 V DC
Max. open voltage Uo UN +25%
Test current IN:
Short-circuit current IK:
Measurement duration:
Overload protection:
Automatic measurement range selection. UUT is automatically discharged after the
measurement.
Low Ohm Measurement (Low Ω)
Measurement Range
0,03...19,99 Ω
Open voltage Uo:
Measurement current:
Measurement duration:
Number of measurements:
Test Lead Compensation
Up to test lead resistance:
Overload protection:
Resolution
1 kΩ
10 kΩ
100 kΩ
1 MΩ
number of measurements with 1mA - Load
(1 measurement - 25 s pause)
approx. 6000
approx. 4000
approx. 3000
approx. 1000
≥ 1 mA at UN
< 2 mA
1 - 3 s
1 - 10 s
optional
10 s, 30 s, 180 s
1200 V AC/DC
Resolution
0,01 Ω
5 V...9 V DC
≥ 200 mA (at resistance ≤ 4 Ω)
approx. 1 s
approx. 3000
max. 1,99 Ω
1200 V AC/DC, as from 600 V AC/DC the internal fuse
trips.
27
Tolerance
±(5% rdg. +10 Digits)
±(5% rdg. +5 Digits)
±(5% rdg. +5 Digits)
±(10% rdg. +5 Digits)
(automatic function)
(automatic function for capacitance loads)
(continuous measurement)
(Timer mode)
Tolerance
±(3% rdg. +3 Digits)
(for 1 Ω, 1 measurement - 25 s pause)