Display Indications

Message
Cause
Reference to exeeding the limit of a measurement value
Interruption / exchange probe(S) - earth probe (ES) (at 4 pole measurement)
Probe
Interruption / exchange probe (S) - earth (E) (at 3 pole measurement)
Interruption auxillary earth (H) - earth (E) (at 2 pole measurement)
Interruption auxillary earth (H) - earth (E) (at 4/3 pole measurement)
Limit
I M
External voltage has exceeded admissible value,
Unoise
Measurement values do not longer correspond to the specifications!
Measurement value RH > 50 kΩ
Limit
or RH > 1000 RE
RH
measurement values do not longer correspond to specifications
Measurement value probe resistance (RS) > 50 kΩ
Limit
or probe resistance (RS) > 10000 x RE
RS
Measurement values do not longer correspond to specifications!
Measurement value earth resistance (RE) > 2 kΩ (at 4/3 pole measurement)
Limit
Displayed measurement values for RH, RS are then RH+RE,RS+RE !
RE
Measurement value R > 2kΩ (at 2 pole measurement)
Limit
Battery-
Batteries are nearly empty. must be changed.
Symbol
Measurement values do not longer correspond to specifications!
Batt-Symbol
Batteries are complete empty. measurement no longer possible!
is
blinking