4 Installation and Commissioning
Damage caused by contaminated and moist cooling air
Contaminated and moist cooling air can cause malfunctions and premature bearing wear.
In general, ensure that a dry, clean and uncontaminated supply of cooling air is provided in accordance with EN ISO 7494-2.
Burning hazard caused by hot turbine
When the turbine overheats, burning in the oral area can occur.
Never contact soft tissue with the turbine!
The user must use the turbine by knowing that an overheating is possible.

5 Operation

5
Operation
Connection to units:
A supply of dry, clean and contamination-free compressed air according to EN ISO 7494-2 must be provided. Contaminated and humid
compressed air will result in premature bearing wear. Screw the turbine coupling onto the tubing and tighten firmly.
Changing the turbine:
To change the turbine, place the turbine straight onto the Connex coupling in an axial direction and push until it clicks audibly into place.
®
Also, compatible with KaVo Multiflex
couplers.
Check for secure seating on the coupling by pulling on the turbine. Inexact coupling, especially during the shutoff delay period, may
destroy the coupler's high-pressure lamp or reduce its service life.
®