1 Important Information
General warnings
To be used by dental practitioners only.
Perform maintenance on a regular basis.
Use good quality tools.
It is recommended that a dental dam be used during treatment.
Never dismantle the turbine. The device may only be dismantled by the approved distributor or at the factory.
Never press the push button while the device is rotating.
It is recommended that suitable protective gloves and eye protection be worn when handling the turbine and instruments.
For extended periods of use, relubricate every 10 minutes.
®
Contact the MICRO-MEGA
after sales department straight away in the event of a problem with the turbine.
Always remove the diamond tip if any part of the diamond is missing.
Check the instrument is present before starting up the turbine.
Warnings regarding reprocessing (cleaning - disinfection - lubrication - sterilisation).
Ensure that the turbine is sterilised and lubricated prior to initial use and between patients.
Remove the rotating parts (burs, files, etc.) before reprocessing.
Avoid soaking the turbine in a tub or ultrasonic cleaning system.
Avoid soaking the turbine in a special purpose detergent for longer than 15 minutes.
Avoid working with a turbine if it is overheating.
Use a detergent-disinfectant designed specifically for pre-disinfection of dynamic dental instruments, which contains neither sodium
oxide nor aldehyde and which is aluminium compatible. Do not use detergents or solutions which are alkaline (pH > 7).
Use the concentration recommended by the pre-disinfection product manufacturer.
Observe the immersion time recommended by the pre-disinfection product manufacturer.
Remove the turbine from the sterilisation bag immediately before use.