3 Reprocessing Methods
Maintenance (Lubrication)
Insert the "500l" spray can into the back of the turbine.
Spray repeatedly for a few seconds each time until the liquid expelled is clean.
Turn the turbine before draining off the excess oil.
For extended periods of use, repeat these steps every 10 minutes.
Do not use vinegar to clean the spray nozzles.
Packaging
Once it has been prepared (manually or automatically) for sterilisation and lubrication, place the turbine in an airtight sterilisation bag:
Individually: A standard packaging material may be used. Ensure that the bag is large enough to hold the instrument without it pulling
on the seal.
In batches: The instruments may be placed in a special purpose instrument carrier or in separate, general purpose sterilisation trays.
Wrap the trays in the appropriate way.
Sterilisation
®
MICRO-MEGA
dynamic instruments can be sterilised in a moist heat autoclave for 18 minutes at between 135° C (275° F) and 138° C
(280.4° F) with pressure of 2.2 bars inside each bag.
Sterilization in a sterilizer compliant with EN 13060
Contact corrosion from moisture can occur
Immediately remove the product from the steam sterilizer after the sterilization cycle is complete!
The turbine has a max. temperature resistance of 138° C (280.4° F).