Maintenance
3.5
After use
1. Uncouple the lung demand valve:
Press
the
button
simultaneously pull the lung demand
valve out of the connector (5).
2. Remove the full face mask:
Pull both mask/helmet adapters to the
rear (6) and flip them away from the
protective headgear.
3. Take off the protective headgear.
4
Maintenance
Following any maintenance work and/or replacement of
components, repeat the leak test.
4.1
Recommended test and maintenance
intervals
The following information complies with DGUV-R 112-190 valid in
Germany. Observe national directives.
Type of
work to be performed
Check by the device user
Visual inspection, function and
leak test
Cleaning and disinfecting
Replace the exhalation valve disc
16
(4)
and
6
4
X
1)
X
X
2)
X
X
X
(approved by Dekra Exam)
Type of
work to be performed
Replace the speech diaphragm
5
1)
Every 2 years for full face masks in airtight packaging.
2)
For masks in airtight packaging; otherwise every 6 months.
4.2
Cleaning and disinfecting
Clean and disinfect the full face mask after every use.
WARNING
!
The mask body must be cleaned sufficiently thoroughly that
no residues of skin care products remain on the mask body.
Otherwise the elasticity in the sealing area can be lost and the
full face mask will no longer sit correctly.
CAUTION
!
Do not use solvents (e.g. acetone, alcohol) or cleaning agents
containing abrasive particles for cleaning and disinfection.
Use only the described procedures and the specified cleaning
agents and disinfectants. Other agents, dosages and contact
times can damage the product.
For information about suitable cleaning agents and
disinfectants and their specifications, refer to document
9100081 at www.draeger.com/IFU.
1. If necessary, remove accessories and clean them separately.
2. Prepare a cleaning solution consisting of water and a cleaning
agent.
X
Panorama Nova S - P