b. Put on belt and close buckle. The device is located on
the back of the user.
c. Tighten belt and fasten protruding ends with clips on
ends of the belt.
8. Connecting the facepiece:
a. Connect the plug-in connector of the breathing hose to
the blower unit.
b. Connect the other end of the breathing hose to the
facepiece.
9. Switch on the blower unit and check the flow rate and
warning devices (see chapter 5.3.4 Check the flow rate
and warning devices)
10. Don the facepiece (see Instructions for Use of the
corresponding facepiece).
11. Adjust the flow rate using the
desired.
3.3
During use
WARNING
Health hazard
► Leave the danger zone immediately in case of:
– Decreasing or interrupted air supply (e.g. after blower
failure). In the hood/helmet/protective visor facepiece
type, carbon dioxide can quickly build up or lack of
oxygen may occur. Noxious ambient air may also
penetrate the hood.
– Odour or taste developing in the facepiece (filter
break). The residual capacity of the gas filter or the gas
filter components of the combination filter are
exhausted.
– Drowsiness, dizziness, or other complaints
– Damage to the equipment
– Displayed alarms
Breathing hoses or other components involve the risk of
getting caught. This may damage the device and interrupt
the air supply!
Handle the device with care.
Breathing in during heavy work while wearing the
hood/helmet/protective visor facepiece type may result in
negative pressure and the penetration of unfiltered
ambient air!
Increase the flow rate to prevent this from happening.
3.3.1 Adjusting the flow rate
If necessary (e.g. during increased physical exertion), the flow
rate must be adjusted during operation using the
buttons.
3.3.2 Warnings and alarms
If a warning appears, leave the working area promptly in view
of the potentially hazardous situation.
Lower the flow rate to increase the period of service if a
warning appears. (Only possible if the lowest level has not
already been chosen.) By lowering the flow rate, you can, for
example, extend the battery runtime.
If an alarm is triggered, leave the working area immediately
without any delay.
Check the function of the device after a warning or alarm has
been triggered.
Instructions for use
|
Dräger X-plore® 8000
and
buttons as
and
3.4
After use
Do the following:
1. Leave the hazardous area.
2. Remove the facepiece (see Instructions for Use of the
corresponding facepiece).
3. Switch off the blower unit by pushing the
control panel for approx. 2 seconds.
4. Open the carrying system belt and take off the device.
5. Clean and disinfect the device (see chapter 5.2 Cleaning
and disinfecting).
4
Troubleshooting
4.1
Warnings
Fault
Cause
The residual run-
time of the
A segment of the
rechargeable bat-
battery status indi-
tery is low
cator is flashing
(< 30 minutes).
yellow.
The particle filter
residual capacity is
A segment of the
low (< 20 %).
particle filter resid-
ual capacity indica-
tor is flashing
yellow.
Malfunction during
switch-on (e.g.
A segment of the
caused by missing
flow rate indicator
hose or filter).
is flashing yellow.
4.2
Alarms
4.2.1 Blower unit
Fault
Cause
The residual run-
time of the
A segment of the
rechargeable bat-
battery status indi-
tery is almost
cator is flashing
exhausted
red.
(< 10 minutes)
The particle filter
residual capacity is
A segment of the
almost exhausted
particle filter resid-
(< 10 %)
ual capacity indica-
tor is flashing red.
Faulty breathing air
supply during oper-
A segment of the
ation (e.g. caused
flow rate indicator
by missing hose,
is flashing red.
filter or kinked
hose).
General system
error
One segment at a
time is flashing red.
|
en
Troubleshooting
button on the
Remedy
Recharge the bat-
tery soon or
replace with fully
charged battery.
Change particle or
combination filter
soon.
Re-check the
device function
and prepare for
use.
Remedy
Recharge the bat-
tery or replace with
fully charged bat-
tery.
Change particle or
combination filter.
Re-check the
device function
and prepare for
use.
Device must be
checked by Dräger
Service.
21