Technical Data - Seca 664 Mode D'emploi

Masquer les pouces Voir aussi pour 664:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 26
... the display flashes?
– If you have just activated the Hold or Tare function, please wait until the scale has de-
tected a steady weight value. The display will then stop flashing.
– If there has been no activation of a function beforehand, remove the load from the scale
0. 0
and wait until
... the display
– The ambient temperature of the scale is too high or too low. Place the scale in an
ambient temperature between +10 °C and +40 °C. Wait about 15 minutes for the sca-
le to adapt to the ambient temperature and then weigh again.
... the display
E
and a number appear?
– Switch off the scale with the Start key and start the scale again. The scale will then
work normally again.
If this is not the case, interrupt the power supply by briefly removing the rechargeable
battery. If this measure is equally unsuccessful, inform the Service department.
9. Maintenance
Your seca scale leaves the factory with an accuracy greater than ± 0.15 %. To ensure that
this degree of accuracy is maintained, the product must be carefully set up and regularly
serviced. Depending on frequency of use, we recommend servicing at intervals of 3 to 5
years. Please note that a qualified servicing agent should be used for this. If you are in doubt,
seca Customer Service will give you more information.

10. Technical data

Dimensions: (width, length, height)
weighing platform:
Scale in upright position:
Weight
Temperature range
Accuracy
Weighing range
Graduations
Colour
Power supply:
24
is displayed, then weigh again.
appears?
800 x 965 x 65 mm
193 x 986 x 1102 mm
approx. 32 kg
+10 °C to +40 °C
greater than ± 0.15 %
0 to 360 kg / 0 to 800 lbs /
0 to 57 sts
100 g / 0.1 lbs
silver grey/black
mains unit and built-in rechargeable
battery pack