Optional Serial Port
The Controller can be read when set in the Front/Remote mode, (read window
NOTE
only enabled), while it can be written to and read when set in the Serial mode
(write window enabled).
Serial Port Installation
An optional RS 485 kit is available for both models. To install it,
proceed as follows:
1
2
3
4
5
6
7
8
9
Turbo-V 3K-T Rack Controller User Manual / 87-900-988-01
Switch off the power and disconnect the power cord.
Unscrew the cover screws and remove the cover.
On rear panel, remove the connector plate, and then secure
connectors J13 to the frame using the turrets provided.
On rear panel, remove the fixing turrets of connectors B and C
and then remove the controller.
Attach the interface board connector to connector P6 on the
controller and then secure it in place using screw A (3x6), which
is not provided.
Insert the flat cable through the rear panel and plug it into the
socket P1 located on the Interface board.
Restore the main board into its original position.
Install and tighten the connector screws and turrets.
Replace the cover.