17
Technical Information
Examples
Window-Based Protocol
The following table, valid for the RS 232, and RS 485 interfaces,
describes each single window used in the protocol.
Tab. 6
WIN
000
001
100
101
102
103
104
107
108
109
116
117
118
119
122
125
126
145
147
167
200
201
202
203
205
230/261
Type
Description
L
START/STOP
L
LOW SPEED [ 0 = OFF / 1 = ON ]
L
SOFT START YES/NO (1= YES ; 0= NO)
L
DELAY EVEN AFTER THRESHOLD [ 0 = OFF / 1 = ON ]
L
WATER COOLING [ 0 = OFF / 1 = ON ]
N
SPEED THRESHOLD (from 0 to 999 Hz)
N
RUN UP TIME (0 359.999 sec)
N
MODE (0, 1, 2) [FRONT, REMOTE, SERIAL]
N
BAUD_RATE (0-4) [600, 1200, 2400, 4800, 9600]
L
PUMP LIFE RESET [write "1" to reset]
N
HIGH SPEED ADJUST [Hz]
N
LOW SPEED ADJUST [Hz]
L
BODY TEMP ENABLE (0=NO, 1=YES)
L
HOST PRINTER MODE [0=HOST / 1=PRINTER]
L
SET VENT VALVE (1=ON ; 0=OFF)
L
SET THE VENT VALVE OPERATION [0=controlled vent 1=on command]
N
VENT VALAVE OPENING DELAY (expressed in sec, max 65535 sec)
L
PURGE OPERATION (0= CLOSE, 1 = OPEN)
N
VENT OPEN TIME (1 sec/bit, 0=always open, max 65535 sec)
L
STOP SPEED READING
N
CURRENT [A]
N
VOLTAGE [V]
N
POWER [W]
N
DRIVING FREQUENCY [Hz]
N
STATUS [0=stop; 1= waiting interlock; 2=starting; 3=normal; 4=high
load; 5=failure; 6=approaching]
Turbo-V 3K-T Rack Controller User Manual / 87-900-988-01