Télécharger Imprimer la page

Error Messages; Error Messages - Agilent Technologies Turbo-V 3K-T Notice De Mode D'emploi

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 37
16
Instructions for Use

Error Messages

Error Messages

For a certain type of failure, the controller will self-diagnose the
error and the messages described in the following table are
displayed.
Tab. 1
MESSAGE
CHECK
CONNECTION TO
PUMP
PUMP WAITING
INTERLOCK
FAULT: PUMP
OVERTEMP.
FAULT:
CONTROLLER
OVERTEMPERATURE
FAULT: TOO HIGH
L0AD
FAULT: SHORT
CIRCUIT
202/260
DESCRIPTION
Wrong connection
between the pump and the
controller.
The interlock signal of P1
connector is activated by
an interruption of the link
between pin 3 and 8 of J1
connector, or because the
external interlock signal is
open.
The pump upper bearing
temperature exceeded 60
°C.
The controller transformer
temperature exceeded 90
°C.
In normal operation, the
current drawn by the pump
is higher than programmed.
In normal operation (after
the starting phase) the
output connection is
shorted.
Turbo-V 3K-T Rack Controller User Manual / 87-900-988-01 (D)
REPAIR ACTION
Check connection between
controller and pump.
Press the START push-button
twice to start the pump.
Reset the short circuit between
pin 3 and pin 8 of J1 connector,
or close the external interlock
signal.
Wait until the temperature
decrease below threshold value.
Press the START push-button
twice to start the pump.
Wait until the temperature
decrease below threshold value.
Press the START push-button
twice to start the pump.
Check that the pump rotor is
free to rotate.
Press the START push-button
twice to start the pump.
Check connections and
shortages between pump and
controller. Press the START
push-button twice to start the
pump.

Publicité

loading

Ce manuel est également adapté pour:

869-9447869-9547