CHARACTERISTICS
Definition:
®
The X-Cid
2 is an automated pre-sterilization cleaning device.
• Ambient temperature for optimal operation: 10°C to 40°C (50°F to 104°F)
• Ambient humidity for optimal operation: 10% to 80% RH
• Atmospheric pressure for optimal operation: 500 hPa (millibars) to 1,060 hPa (millibars).
Dimensions:
H=47cm /L=40cm /W=23cm
Weight:
8 kg
Electrical:
Ref.: 51100003 = 230V +/- 10% 50-60 Hz EU
Electric class: I
Power:
38 VA
Protection:
• Two 0.2 Amp slow-blow fuses (5x20 mm), Ref.: 50900014, for the 230 V X-Cid
®
2
• Two 0.5 Amp slow-blow fuses (5x20 mm), Ref.: 50900015, for the 100 and 110 V X-Cid
®
2
Electrical safety:
The device complies with the following standards:
• EN 60601-1-2 : 2007 Electromagnetic compatibility.
• EN 61010-1 : 2001 Safety requirements for electrical equipment for measurement, control and laboratory.
• EN 61010-2-040 :2005 Particular requirements for sterilisers and washer-disinfectors used to treat medical equipment.
Compressed air feed:
• Feed pressure: max. = 8 bars / min. = 5 bars
• Average flow: 2 l/minute
• Air quality: 1 micron filter; residual water/oil=0.5 ppm at 25°C
• Connection: quick-connect for 4x6 tubing
Evacuation:
Tube (Ø 25 mm ext.) connected directly to the waste fluid outlet without siphon.
Treatment fluids:
• Pre-disinfection and cleaning: X-Cidol
®
2 detergent; Ref.: 50704005
• Rinsing: Demineralised water with conductivity between 0.5 and 10 µS/cm, do not use scented water.
• Lubricating: MICRO-MEGA
®
TSU Oil Ref.: 50700025
25
Année du marquage CE : 2006
07/13 Ref. 60300544-I
Year of CE-marking: 2006