GENERAL INFORMATION
Introduction
■
®
The X-Cid
2 automatically pre-disinfects, cleans and lubricates handpieces, contra-angles, and turbines with no prior
disassembly required.
■
The treated instruments are ready to be packaged and sterilised.
■
®
Since the X-Cid
2 complies with Administrative Circular N° DGS/5C/DHOS/E2/2001/138 of 14 March 2001, patients and
health personnel can be assured that there is no risk of cross-infection. In addition, the X-Cid
instruments, reduces maintenance costs, and is compatible with all brands of handpieces, contra-angles, and turbines.
PROCEDURE
Pre-disinfection (15 min)
■
All surfaces in contact with pathogens are treated with X-Cidol
complies with current standards. The internal parts of instruments rotate at low speed to eliminate interfering substances as
well as most microorganisms.
Rinsing (2 min 15 s)
■
The rotating instruments are rinsed with demineralised water to evacuate the X-Cidol
Cleaning (5 min 30 s)
■
The instruments rotate again in X-Cidol
microorganisms, leaving the instruments totally clean.
Rinsing (2 min 15 s)
■
The rotating instruments are rinsed with demineralised water to evacuate the X-Cidol
Drying (2 min)
■
Once clean and rinsed, the instruments are dried using compressed air.
Lubricating (4 min 15 s)
■
The rotating instruments are then lubricated using special oil applied using a mounted spray can.
The excess lubricant is removed using compressed air.
Année du marquage CE : 2006
Year of CE-marking: 2006
CE-Kennzeichnung: 2006
®
2, an aldehyde-free detergent/ disinfectant solution that
®
2. This phase complements the pre-disinfection phase and eliminates all remaining
®
2 increases the service life of
®
2 used in the previous phase.
®
2 used in the previous phase.
07/13 Ref. 60300544-I