GENERAL INFORMATION

Introduction
®
The X-Cid
2 automatically pre-disinfects, cleans and lubricates handpieces, contra-angles, and turbines with no prior
disassembly required.
The treated instruments are ready to be packaged and sterilised.
®
Since the X-Cid
2 complies with Administrative Circular N° DGS/5C/DHOS/E2/2001/138 of 14 March 2001, patients and
health personnel can be assured that there is no risk of cross-infection. In addition, the X-Cid
instruments, reduces maintenance costs, and is compatible with all brands of handpieces, contra-angles, and turbines.
PROCEDURE
Pre-disinfection (15 min)
All surfaces in contact with pathogens are treated with X-Cidol
complies with current standards. The internal parts of instruments rotate at low speed to eliminate interfering substances as
well as most microorganisms.
Rinsing (2 min 15 s)
The rotating instruments are rinsed with demineralised water to evacuate the X-Cidol
Cleaning (5 min 30 s)
The instruments rotate again in X-Cidol
microorganisms, leaving the instruments totally clean.
Rinsing (2 min 15 s)
The rotating instruments are rinsed with demineralised water to evacuate the X-Cidol
Drying (2 min)
Once clean and rinsed, the instruments are dried using compressed air.
Lubricating (4 min 15 s)
The rotating instruments are then lubricated using special oil applied using a mounted spray can.
The excess lubricant is removed using compressed air.
Année du marquage CE : 2006
Year of CE-marking: 2006
CE-Kennzeichnung: 2006
®
2, an aldehyde-free detergent/ disinfectant solution that
®
2. This phase complements the pre-disinfection phase and eliminates all remaining
®
2 increases the service life of
®
2 used in the previous phase.
®
2 used in the previous phase.