Daily maintenance

Before the first daily use of the X-Cid
automatically and an alarm will be set off when fluid levels are insufficient.

Monthly maintenance

1) Make sure that the lubricating system works correctly by running a lubricating cycle without cleaning (follow instructions
hereafter):
Before starting the maintenance procedure, remove all the rotating instrument holders and the turbine couplings.
Close the treatment tank door.
- Press the button "1" for 5 seconds: Once the yellow LED "1" remains lit, release the button and instantly
- press the button "2" for 5 seconds: Once the yellow LED "2" remains lit, release the button and instantly
- press the button "3" for 5 seconds: Once the yellow LED "3" remains lit, release the button.
Open the treatment tank door and place a sheet of paper under the three stations. After two minutes, oil marks should
appear under each station (see photo).
Then close the treatment tank door and wait until the end of the drying cycle (3 min). The device will stop automatically and
the three LEDs 1-2-3 will blink.
2) Select all three stations, operate the device without instrument, and verify that there is a good flow of fluid at each station.
Then interrupt the cycle by pressing the "power on" button.
3) Clean the outside of the device with a soft towel dipped in disinfectant. Never use abrasives. Never spray any cleaning
product on the keypad.
4) Brush the drainage grid at the bottom of the tank (see photo hereafter).
Then close the treatment tank door after having reconnected your clean turbine coupling(s).
5) Clean the reservoir for the X-Cidol
6) Clean the removable reservoir under running water, remove the filter and clean it with a soft brush.
7) Rinse the device following the instructions of the maintenance and control sheet, Ref.: 60300552. Make sure that the X-
®
Cidol
2 reservoir is nearly empty before starting the procedure.
8) The procedure is now completed and the device is ready for use.
Fuses
■ The fuses are located near the ON/OFF switch. They must be replaced by fuses of the same type (see characteristics).
Année du marquage CE : 2006
Year of CE-marking: 2006
CE-Kennzeichnung: 2006
®
2, verify the fluid levels. In case you forget this step, the cycle will stop
®
2 with a disinfectant, non pilling wipe without removing the filter.