2) Treatment tank:
■
The treatment tank, which has a window, has three stations that can
accommodate instrument couplings in any order. Each coupling is held in
place by two screws (1). Two types of couplings can be used:
• an ISO 3964 pivot coupling for handpieces and contra-angles (2);
• an ISO 9168 (MID) pivot coupling for turbines (3). These couplings can be
equipped with quick-connectors for each turbine.
■
The order of the coupling can be changed to accommodate the types of
instruments being used. It is recommended to maintain the MID coupling in
a central position.
■
Each station is equipped with four removable spray nozzles to treat the
external surfaces of the instruments (4).
After treatment of the contra-angles, press the button (5) above the coupling
in order to disconnect the contra-angles from the device.
5
2) Product reservoir:
■
The reservoir is located in the top of the device and is composed of two compartments with independent hinged lids. It has
been designed for use with the following products:
• X-Cidol
®
2 detergent / disinfect for the pre-disinfection and cleaning phases.
• Demineralised water for the rinsing phases.
■
To fill, the covers of the two compartments are opened and the appropriate quantity of product is added up to the fill line
using a funnel, if necessary. Do not exceed the maximum fill line in order to avoid overflows. In case of liquid spillage: Wipe
the spilled liquid immediately.
■
Cleaning: The water reservoir can be removed by pulling vertically on its strap (photo B), its filter can also be removed.
After removal, clean the filter by brushing it under running water (photo C). It is recommended only to fill the water reservoir
once it has been fixed onto the device. For fixing it on the device, press on the filter (photo D). Make sure that the X-Cidol
reservoir is perfectly clean. Do not remove the filter in case of product residues (photo E).