• Only select those stations with an attached instrument. Otherwise, cleaning product will be wasted.
• Make sure the stations to be used are correctly selected on the keypad.
6. Press the "start cycle" button.
• During the cycle, the yellow "cycle" and "stations in use" LEDs will light up.
• At the end of a cycle, the "cycle" LED will turn off and the "stations in use" LEDs will blink.
• If the user voluntarily stops the cycle, the "stations in use" LEDs will turn off (which indicates a reset). The user must
reselect the stations before starting a new cycle.
7) At the end of every cycle: Open the hinged door and remove the instruments from the treatment tank. Do not leave the
instruments overnight or during the week-end in the device.
Lubrication modes
You can operate your X-Cid
time as the "power on" button.
Button 1 + "Power on" button: Non-lubricating mode
Button 2 + "Power on" button: Standard lubricating mode
Button 3 + "Power on" button: Intensified lubricating mode
The lubricating mode will be displayed and the red LED will light up for 5 seconds when the power is turned on.
IMPORTANT INFORMATION: At the end of a non-lubricating mode cycle, the device will automatically switch back to the
standard lubricating mode.
MAINTENANCE AND RECOMMENDATIONS
®
The X-Cid
2 has been designed to clean and lubricate handpieces, contra-angles and turbines after each patient. The
following recommendations will help you make the best use of the device and your instruments.
■ If your instruments are not new, we recommend a complete internal cleaning to remove accumulated debris prior to a
®
treatment with the X-Cid
2 or to make 2 or 3 cycles.
■ Make sure that the internal canals of your contra-angles are wide enough for the cleaning liquids to pass through.
■
Treat your instruments immediately after use so that saliva and blood do not have time to coagulate or dry.
■
Once treated, instruments must be packaged for sterilization.
■
During long dental procedures (> 15 min) with fast rotating instruments (contra-angles with red rings or turbines), we
recommend that you manually lubricate your instruments during the procedure because the air flow tends to eliminate the
lubricant.
■ If the X-Cid
®
2 is not used for a longer period of time, rinse the device following the instructions of the procedure sheet ref.
60300552.
■ We recommend annual preventive service by a manufacturer-certified technician. This control includes liquid volume
measurement, sensor functioning, alarms, lubrication, change of filters and reservoir joints if required.
IMPORTANT INFORMATION: For cleaning and maintenance of your X-CID
products that contain chlorine (bleach) or hydrochloric acid. Products that contain these ingredients create brown stains on
stainless steel parts which can deteriorate your device and your rotating instrument holders.
Année du marquage CE : 2006
Year of CE-marking: 2006
CE-Kennzeichnung: 2006
®
2 in three lubrication modes by pressing one of the buttons on the keypad (1, 2 or 3) at the same
®
2 device, never use cleaning wipes or other
07/13 Ref. 60300544-I