X-Cid
2
®
USER MANUAL
This device has been designed and manufactured in accordance with the European directive 94/42 CE (class I).
®
X-Cid
2 is an automated device for cleaning dental handpieces, contra-angles and turbines.
LISTE DES SYMBOLES
SAFETY INSTRUCTIONS
■ The X-Cid
®
2 must only be used by qualified personnel to pre-disinfect handpieces, contra-angles and turbines.
■ The instruments must be in good working order. It may not be possible to properly clean a damaged instrument.
■ Always wear clean, disposable medical gloves when handling handpieces, contra-angles and turbines and X-Cidol
■ Only use manufacturer recommended products with the X-Cid
• X-Cidol
®
2 detergent; Ref.: 50704005
• Demineralised water
• TSU Oil MICRO-MEGA
®
• The device must be installed by a manufacturer-certified technician following the instructions in the installation manual.
■ Caution when filling the X-Cidol
your eyes and face. The risk for severe eye damage is high.
■ Do not inhale: Inhalation of steams may cause drowsiness and vertigo. In case of eye contact, wash off immediately and
profusely with water and refer to a specialist.
■ For more details: please report to the X-Cidol
■ In case of heavy overflowing, when filling the X-Cidol
again.
■ Do not open the device during its use.
■ Following an abnormal functioning, and in case the scuttle lid should open, the wear of a surgical mask and gloves is
recommended before attempting to stop the device.
■ In case the locking system should be damaged, please refer to the salesman.
■
Do not store the device near a heat source.
■
®
Do not expose the X-Cid
■
Do not smoke in close proximity to the device.
■
The air exchange rate in the room where the device is installed must be at least 100m
■
If the device is not used for four weeks, complete an empty cycle before installing instruments in the treatment tank.
■
If detergent comes in contact with the eyes, rinse immediately and thoroughly with water and consult a physician. Consult
®
the X-Cidol
2 safety data sheet available from your distributor or MICRO-MEGA
■
Do not disassemble or damage the device.
■
No not place your hands on the hinges when opening the window.
■
Make sure the device is solidly seated before use.
■
The wall socket must be grounded.
■
To unplug the device, remove the power cable.
■
Unplug the device and contact your distributor:
• If the device seems to be operating abnormally (time cycles not respected, abnormal liquid consumption, abnormal noise,
etc.).
• If the liquid is no longer evacuated from the reservoir.
• If there is a leak at the back of the device.
• If there is visible damage to the device.
■
Contact your distributor or MICRO-MEGA
with EU Directive 2002/96/CE. .
Année du marquage CE : 2006
Year of CE-marking: 2006
CE-Kennzeichnung: 2006
Fragile package contents
Safety instructions
Recycling process for end-
of-use produtcs
; Ref.: 50700025
®
2 tank: You must wear an adequate protective gown, gloves, and a protective device for
®
2 Security Data File, delivered with the device.
2 to direct sunlight.
®
in order to dispose of the device at the end of its service life in accordance
®
2:
®
2 tank, leave the device to dry out for 24 hours before using it
3
/h.
®
.
CE Marking
Manufacturer
®
2.
07/13 Ref. 60300544-I
16