3. QC Mode
Select QC mode to set up the analyzer to act as a quality control device,
which will enable you to test up to 200 batteries in succession without
resetting test parameters. (You can change parameters at any time
during QC testing. The results will be displayed or printed along with the
new parameters.)
1. On/Off: If you turn on QC mode, it will stay on until you turn it off.
The analyzer will go into Sleep mode to save the 9-volt battery power
after 30 continuous seconds of voltage loss at the clamps. When
voltage is restored, the analyzer will continue saving data in QC mode
to the next memory location.
2. Clear Memory, Yes/No: The analyzer's memory holds up to 200 test
results in QC memory. Select YES to clear test results from memory
or NO to continue storing results. When the memory is full, the
analyzer will display the following warning before it starts to overwrite
earlier test results:
MEMORY IS FULL / PRESS ↵ TO OVERWRITE NO. 1
If you want to export or print the stored results before clearing the
memory, press MENU to exit.
3. Select Input/Test By: Select a rating system to TEST BY: CCA, CA,
MCA, or DIN.
4. Set Minimum Amps: Select the CCA threshold for Pass/Fail.
5. Set Minimum Voltage: Select the voltage threshold for Pass/Fail.
6. To start QC mode testing, connect the analyzer and test each battery
until finished. Press the MENU button to return to the Options Menu.
4. Voltmeter
The analyzer will function as a voltmeter (with an operating range of
0 through 30 Vdc) when you first connect it to a battery.
You can turn this feature on or off, depending on your preference. Press
the MENU button; then select the Voltmeter option. Press ENTER, then
turn the voltmeter ON or OFF.
E
NGLISH
15
•
•