Final test values related to the nominal capacity (maximum volume) indi-
cated on the instrument, obtained when instrument and distilled water
are equilibrated at ambient temperature (20 °C/68 °F) and with smooth
operation. According to DIN EN ISO 8655.
Accuracy tolerances for the Transferpette
Volume range
Volume step
µl
µl
500 - 5000
5000
2500
100 - 1000
1000
20 - 200
2 -
20
0.5 -
10
A* = Accuracy, CV** = Coefficient of Variation
Note:
Testing instructions (SOPs) and a demo version of the EASYCAL
available for download at www.brand.de.
®
A*
≤ ± %
0.6
1.0
500
3.0
0.6
500
1.0
100
3.0
200
0.8
100
1.2
20
4.0
20
1.0
10
1.5
2
5.0
10
1.0
5
1.5
1
5.0
electronic
CV**
≤ %
0.2
0.3
0.6
0.2
0.3
0.6
0.2
0.3
0.6
0.4
0.8
2.5
0.4
0.8
2.0
4.0 calibration software are

Accuracy Table

20 °C
!
Ex
Increment
Type of tip
µl
µl
5.0
5000
1.0
1000
0.2
200
0.02
20
0.01
20
55