b) Results Detail page
Status
History
Print
Result
Start Time
Well# and Configuration
390 Serial Number
Control?
BI Lot#
Load#
Sterilizer#
Technician
Chemical Integrator?
Implant?
Cycle Parameters
Notes
This page provides an individual record for an incubated BI. For complete documentation,
the user may print a hard copy and/or save the record as an electronic file.
Note 1: The top four rows (highlighted in green) are populated automatically by the
3M™ Attest™ Auto-reader 390 and cannot be changed by the user. The remaining rows
are completed by the user to satisfy the documentation requirements of their facility.
Note 2: The Results Detail page refreshes less frequently that the Status page. Until the BI
result is available, the Results Detail page displays remaining minutes of incubation in the
results field.
c) History page
Status
ID
Well# and Configuration
History
1
#9 1292
2
#4 1291
3
#8 1292
4
#3 1291
5
#7 1292
6
#2 1291
7
#1 1291
8
#6 1292
The 3M™ Attest™ Auto-reader 390 stores results for the last 100 incubated BIs on
embedded software. The 3M™ Attest™ Auto-reader 390 Web App History page provides
a summary of these last 100 results.
d) Configuration page
This page allows the configuration of the 3M™ Attest™ Auto-reader 390's network and
language settings from the 3M™ Attest™ Auto-reader 390 Web App.
Documenting BI Results using the 3M™ Attest™
Auto-reader 390 Web App
Status
History
ATTEST 1291
1
2
3
4
+
59
–
–
1. Open the Results Detail page
After inserting a new 3M™ Attest™ Rapid Readout BI in the 3M™ Attest™
Auto-reader 390, it is a best practice to open the 3M™ Attest™ Auto-reader 390 Web
App Results Detail page for that BI by clicking on the appropriate well number from
the Status page.
10
3M™ Attest™ Auto-reader 390 Results Detail
+
2018-10-18 08:08:39
#6 1292
200004
Yes
2015-02TI
n/a
n/a
sdf
n/a
n/a
n/a
Daily control BI
3M™ Attest™ Auto-reader 390
Start Time
Result
2018-10-18 14:31:33
---
2018-10-18 14:23:10
---
2018-10-18 11:17:04
–
2018-10-18 11:16:59
–
2018-10-18 09:10:03
–
2018-10-18 09:10:02
–
2018-10-18 08:09:31
+
2018-10-18 08:08:39
+
3M™ Attest™ Auto-reader 390
ATTEST 1292
5
6
7
8
9
+
–
–
174
2018-10-18 15:27
SN
200004
2. Enter the specific information for that BI (e.g., BI Lot #, sterilizer/load number,
Technician name, etc.) in the provided editable fields. Use the Notes section to capture
information not prescribed in the provided row titles; for example, you may wish to use
this field to record load items.
Status
History
Print
Result
Start Time
Well# and Configuration
390 Serial Number
Control?
BI Lot#
Load#
Sterilizer#
Technician
Chemical Integrator?
Implant?
Cycle Parameters
3. The entered information is saved automatically to temporary web browser storage on
the user's PC every 60 seconds, or by leaving the Results Details page.
4. Print/File. Once the BI result is available, print or file the Results Detail page to create
a record. The page may be physically printed and/or saved as an electronic file.
Important! Data is only stored for the last 100 incubated BIs.
Embedded software in the 3M™ Attest™ Auto-reader 390 saves the data from the
auto-populated (i.e., green highlighted) fields. The application relies on the temporary
storage in the user's web browser to save the data in the user-populated fields. In both
cases, data is only stored for the last 100 records, i.e., the 101st incubated BI will erase
the information saved for the 1st incubated BI. To retain the Results Detail information for
future reference it is best practice to print the page or save to an electronic file as soon as
the BI result is available.
Using the BI History Page
Status
ID
Well# and Configuration
History
1
#9 1292
2
#4 1291
3
#8 1292
#3 1291
4
5
#7 1292
6
#2 1291
7
#1 1291
8
#6 1292
To view or print an individual record, click on the ID number of the desired record to
navigate to the Results Detail page for that BI.
Note 1: Column 1, "ID" is the assigned ID number with ID 1 representing the newest
record and ID 100 representing the oldest record. For convenience, all columns are
sortable; to help find a particular record, simply click once on the column heading to sort
in ascending order and click a second time to reverse the order.
Note 2: The History page refreshes less frequently than the Status page. Until the BI result
is available, the History page displays ••• in the Results column indicating incubation is
still in-process.
10
3M™ Attest™ Auto-reader 390 Results Detail
54
2018-10-18 15:26:42
#2 1291
200004
No
2015-01DS
1
1
sdf
accept
no
270F 4min
3M™ Attest™ Auto-reader 390
Start Time
Result
2018-10-18 14:31:33
---
2018-10-18 14:23:10
---
2018-10-18 11:17:04
–
2018-10-18 11:16:59
–
2018-10-18 09:10:03
–
2018-10-18 09:10:02
–
2018-10-18 08:09:31
+
2018-10-18 08:08:39
+