TC-Standard (+)
Gas cooler
(two single heat exchangers)
Heat exchanger 1
Heat exchanger 2
gas in
gas in
gas out
gas out
condensate out
condensate out
The gas inputs are marked in red.
On glass heat exchangers the correct position of the seal is im-
portant when connecting the gas lines (see image). The seal
consists of a silicone ring with a PTFE sleeve. The PTFE side
must face the glass thread.
Glass
PTFE

4.3 Electrical connections

The operator must install an external separator for the device
which is clearly assigned to this device.
This separator
• must be located near the device,
• must be easy for the operator to reach,
• must comply with IEC 60947-1 and IEC 60947-3,
• must separate all live conductors and the status output,
and
• must not be attached to the power feed.
The mains supply of the device must be fused according to the
specifications under technical data.
Potential equalization/static charge
Static charges can result in incendive sparking. Avoid static
charges. Any conductive parts of the cooler must be groun-
ded!
The housing has a connection for an earth/equipotential
bonding conductor. Ensure the housing is adequately earthed
(minimum conductor cross-section 4 mm
WARNING
Hazardous electrical voltage
The device must be installed by trained staff only.
12
Bühler Technologies GmbH
Gas cooler +
(two heat exchangers in series)
Heat exchanger 1
Heat exchanger 2
gas out
gas out
gas in
gas in
condensate out
condensate out
2
).
CAUTION
Wrong mains voltage
Wrong mains voltage may damage the device.
Regard the correct mains voltage as given on the type plate.
WARNING
High voltage
Damage to the device in case of insulation testing
Do not proceed insulation tests with high voltage to the
device as a whole!
Electric strength test
This device is equipped with extensive EMC protection. The
necessary tests were carried out at the factory (test voltage
2.1 kV or 2.55 kV DC depending on approval).
If you wish to check the electric strength again yourself, you
can do so on the entire unit. Only test the device with the spe-
cified values using direct current voltage. Testing the electric
strength with alternating current voltage will damage elec-
tronic components. The recommended voltage in this case is
2.1 kV DC, 2 s. Disconnect all supply lines from the device be-
fore testing. Power can be supplied via the mains connection.
Connection via plug
The device is equipped with connectors according to EN
175301-803 for mains and status output. If the cables are
mounted properly, they cannot be interchanged. Please make
sure that the connectors are remounted correctly after con-
necting the cable. The following figures show the pin assign-
ment with respect to the numbers printed on the connector.
The supply line cross-sections must be suitable for the rated
current. Use a maximum line cross-section of 1.5 mm²
(AWG 16) and a cable diameter of 8 - 10 mm (0.31 - 0.39 inch).
Plug numbering
Alarm contact S2 *
*
* When using an alarm contact, connect to ≥ 33 V AC or
≥ 70 V DC PE voltage.
The clamping area is 8 - 10 mm (0.31 – 0.39 inches) in dia-
meter.
Electric supply S1
115/230 V (24 V DC)
(+24 V)
(-24 V)
Analogue output S3
1 - N/C
2 - N/C
3 - GND
4 - 4-20 mA out
5 - shield
BX440026 ◦ 12/2020