Setup Procedures
English
Unpack the cartons and check all items for proper operation. It is important that the Stryker Evacuation Chair is
working properly before it is put into service. Have a qualified service person use the following list and the operation
instructions to check the chair before it is put into service.
• All fasteners secure (reference all assembly drawings).
• All welds intact, not cracked or broken.
• No bent or broken tubing or sheet metal.
• No debris in wheels.
• All wheels secure and rolling properly.
• Chair unfolds and locks properly.
• No rips or cracks in seat or backrest.
• Passenger restraints intact and working properly.
• Wheel locks operating properly.
• Foot end lift handles extend and lock properly.
• Head end lift handles fold and unfold.
• Casters secure, rolling and swiveling properly
• Upper control handle extends and locks in both positions
• Stair-TREAD™ system unfolds and locks
• Track belts roll properly
• Optional accessories intact and operating properly.
WARNING
Do not modify the Stryker Evacuation Chair. Modifying the chair can cause unpredictable operation resulting in injury to
the passenger or operator. Modifying the chair will also void its warranty.