The filter is produced with (PTFE) hydrophobic material witch prevents fluids entering the pneumatic circuit.
When the filter is wet, it's not possible to use the unit therefore the filter should be changed immediately.
In case of possible contamination or discolouration, change the filter immediately.
Don't use the suction unit without the protection filter fitted. If the suction unit is used in an emergency or in a
patient where the risk of contamination is not know the filter must be changed after each use.
Available under request with different versions with complete jar 2000cc.
The suction probe must be changed at the end of each application.
Wearing disposable latex gloves, proceed to clean the accessories as follows:
1.
Grip the jar with your hand and turn the cap in a counter-clockwise direction.
2.
Remove the red cage covering the float located in the jar cap (conical part with 3 openings)
3.
Once the cage is removed, the float with the red seal comes out of the jar cap.
4.
To disinfect the accessories, use either denatured alcohol or hypochlorate-based solutions,
easily purchasable from chemist shops.
5.
Once cleaning has been completed, insert the float in the cage, with the seal facing towards the cage
opening.
6.
Insert the opening of the cage into the seat in the jar cap.
7.
To facilitate the cleaning of the jar, fill it 1/3 full of ordinary water. The aspirated liquid, thus diluted, will
then be easier to remove.
It is possible autoclave the accessories cover and jar: insert the parts in autoclave and to effect a cycle of
sterilization having care to position turned upside-down the graduated jar (with turned fund toward the tall
one). After the sterilization and the cooling to the temperature the components' environment, to verify that this
last don't result damaged; disassembled therefore the container for inhaled liquids following the inverse
operations to the dismantlement.
The aspiration tubes can be sterilized on autoclave using a sterilization cycle at 120°C.
The conical connector can be sterilized on autoclave using a sterilization cycle at 121°C.
The device is ready for a new employment now.
During the sterilization operations the operator must use disposable gloves in latex, shirts and opportune
mask not to enter contact with possible substances contaminating
DO NOT WASH, STERILIZE OR PUT IN AUTOCLAVE THE ANTIBACTERIAL FILTER

ACCESSORIES SUPPLIED

DESCRIPTION
COMPLETE ASPIRATION JAR 1000cc
CONICAL FITTING
TUBES SET 6 mm x 10 mm
ASPIRATION PROBE CH20
ANTIBACTERIAL FILTER

CLEANING OF ACCESSORIES