Mettler Toledo SevenGo pro SG7 Mode D'emploi page 20

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 53
Menus
GLP
There are 2 data formats for your selection when outputting the data from memory: GLP or Non-GLP:
Examples:
GLP format
GLP
On
Date
01-JAN-05 / JAN-01-05
Time
09:31:03
Sample_ID
000326
Result
20.7mS/cm
Temperature
25.3 °C
Ref.Temp.
25 °C
Temp.Comp.
2.0%/°C
ATC/MTC
ATC
Endpoint
Auto
Sensor_ID
04102601
Last_cal.
10-23-04
Signature
________
Non-GLP format
GLP
Off
Result
20.7mS/cm
Temperature
25.3 °C
Ref.Temp.
25 °C
Temp.Comp.
2.0%/°C
ATC/MTC
ATC
Endpoint
Auto
System
"Light off"
You can define the period (10 s, 15 s, 30 s) after which the back lighting switches of automatically.
Note
The light function can be turned on/off by just pressing and holding the
"Self test"
This menu item starts the self diagnosis routine. The meter displays the full screen first, then each icon blinks
one after the other. This way you may check whether all icons are shown regularly. The final step is to check
that the keys are functioning correctly. This requires user interaction.
You are now requested to press the seven function keys on the keypad one by one in any order: after press-
ing a key, an icon disappears from the display; continue to press the other keys until all the icons have
disappeared.
When self diagnosis has been terminated succesfully the meter returns to the system menu. If errors are
reported contact METTLER TOLEDO Service.
Note
You have to finish pressing all the seven keys within two minutes, otherwise "Self test failed!" appears, and
you will have to repeat the procedure.
16
button.
METTLER TOLEDO SevenGo pro™ conductivity

Publicité

Table des Matières
loading

Table des Matières