IKA LR-2.ST Mode D'emploi page 16

Masquer les pouces Voir aussi pour LR-2.ST:
Table des Matières

Publicité

The maximum permitted temperature inside the reac-
tor vessel is 230 °C. Do not exceed this temperature.
Before you fill the reactor vessel, ensure that the
reagents used do not corrode the seal!
Accessories
Only use IKA
approved accessories!
®
Use only original IKA
Protect the instrument and accessories from bumping
and impacting.
Please observe the permitted speed for the stirring
element. Never set higher speed.
Assembly
Ensure that the stirring tool is
WARNING
securely fixed.
Broken glass! Fit the hose cou-
CAUTION
pling carefully.
• Fix all hoses in the laboratory extractor hood.
Ensure that the glass assembly is tension-free! Danger
of cracking as a result of:
- stress due to incorrect assembly,
- external mechanical hazards,
- local temperature peaks.
16
spare parts!
®
Stand
Undo the screws on the stand
CAUTION
carefully.
balance the stand can move
upwards uncontrollably.
Protection of the device
The voltage stated on the type plate must correspond
to the mains voltage.
Socket must be earthed (protective ground contact).
supply.
Covering or parts that are capable of being removed
from the unit without accessory equipment have to be
reattached to the unit for safe operation in order to
prevent, for example, the ingress of fluids, foreign mat-
ter, etc..
The appliance may only be opened by experts.
Without
counter

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

20000005261

Table des Matières