- Connect the power cord to the receptacle of the
device.
- Connect the power cord to a power socket with a
ground pin.
- If necessary, connect the potential equilization
cable of your installation system to the unit
potential equalization connector (Fig. 2-6)
- Connect the control footswitch to the footswitch
connector (Fig. 2-2).
- Position the footswitch so that it is easily accessible
for your feet.
- Insert the brackets into the bracket holders (Fig. 2-
4).
- Connect the micro-motor cord to the connector
(Fig. 1-7).
- Align the I-S
URGE
connector electrical contacts, then screw the motor
onto the connector.
- Connect the N
EWTRON
2 L
handpiece cord to the connector (Fig. 1-9)
ED
(according to option).
- Connect the N
EWTRON
cord (delivered according to option).
- Check that the device is close enough to the work
area to be used without pulling on the cords. If it is
not, move the device closer.
- Place the I-S
L
URGE
ED
- Place the N
L
EWTRON
on the handpiece support.
- Suspend the bottles or bags of physiological saline
or sterile water from the brackets.
- Adjust the position of your device to suit your angle
of view.

4. 4 FIRST USE

Before using the I
MPLANT
the equipment must be given routine maintenance
and/or sterilized according to the procedures defined
in chapter 11.
L
micro-motor and cord
ED
L
scaler cord or the P
ED
L
handpiece to the scaler
ED
micro-motor on its rest.
or P
2 L
ED
IEZOTOME
ED
C
2 for the first time, all
ENTER
V - IRRIGATION SOLUTION
I
C
MPLANT
must be used only with bottles or bags of
physiological saline or sterile water with an individual
capacity not exceeding one litre.
VI - IRRIGATION LINES
I
MPLANT
SATELEC irrigation lines.
VII - ADJUSTMENTS /
IEZOTOME

7. 1 STARTUP

Important:
After stopping the device, wait four seconds before
restarting it.
handpiece
Each time the device is switched on, the screen
displays program P1 of the I-S
displaying the welcome page.
7. 2 SETTINGS
The settings are saved when each program is
configured and are retrieved whenever that program
is selected.

7. 3 IRRIGATION FUNCTION

a) Flush / prime
The flush / prime function is available on the LCD
screen and on the footswitch.
2 is not designed to administer drugs. It
ENTER
C
2 must be used exclusively with
ENTER
SETTINGS / MODES /
INTERFACE
function after
URGE
11