e) Control footswitch
Having access to a large number of functions on the
control footswitch allows the practitioner to work in
a perfectly sterile environment, avoiding the risk of
cross-contamination.
Once the various settings have been adjusted, the
user has no further need to touch the keypad.
Depending on the mode, the control footswitch
buttons have different functions.
I-S
mode:
URGE
Footswitch button definitions (Fig. 11):
- 1: I-S
L
motor control (ON/OFF or
URGE
ED
progressive).
- 2: Rotation direction.
- 3: Change program (P1 to P4).
- 4: Irrigation ON/OFF.
- 5: Select active handpiece.
P
mode:
IEZOTOME
Footswitch button definition (Fig. 11):
- 1: Ultrasonics control (ON/OFF or progressive).
- 2: Flush/prime.
- 3: Change program (D1 to D4).
- 4: Irrigation ON/OFF.
- 5: Select active handpiece.
N
mode:
EWTRON
Footswitch button definition (Fig. 11):
- 1: Ultrasonics control (ON/OFF or progressive).
- 2: Flush/prime.
- 3: Change program (Soft to Boost).
- 4: Irrigation ON/OFF.
- 5: Select active handpiece.
f) Technical characteristics
Manufacturer: SATELEC
Device name: I
MPLANT
Electrical power supply:
- Voltage: 100 VAC to 230 VAC.
- Frequency: 50 Hz/60 Hz.
- Rated power: 250 VA at 230 VAC.
C
2
ENTER
I-S
function
URGE
Operation:
Intermittent service: 20 sec. ON / 30 sec. OFF at 2
N.cm
Output characteristics:
I-S
L
micro-motor speed: 100 rpm to
URGE
ED
40 000 rpm
I-S
L
micro-motor torque: max. 6 N.cm
URGE
ED
Irrigation flow rate: 10 to 120 ml/min. (nominal
value).
Adjustment in 10 ml/min. steps
Flush/prime flow rate: 120 ml/min.
P
function
IEZOTOME
Operation:
Intermittent service: 10 min. ON / 5 min OFF
Output characteristics:
No-load voltage: 250 volts (nominal value without
handpiece)
Min. ultrasonic frequency: 28 kHz
Irrigation flow rate: 10 to 120 ml/min. (nominal
value)
Adjustment in 10 ml/min. steps (nominal value)
Flush flow rate: 120 ml/min. (nominal value)
N
function
EWTRON
Operation:
Intermittent service: 10 min. ON / 5 min. OFF
Output characteristics:
No-load voltage: 150 volts (nominal value without
handpiece)
Min. ultrasonic frequency: 28 kHz
Irrigation flow rate: 10 to 40 ml/min. (nominal value)
Adjustment in 1 ml/min. steps (nominal value)
Flush flow rate: 120 ml/min. (nominal value)
Protection
Electrical equipment class: Class 1
Electrical safety class:
Type BF in I-S
mode
URGE
Type BF in P
mode
IEZOTOME
Type BF in N
mode
EWTRON
9