1. 2 TECHNICAL DESCRIPTION

a) Front panel (Fig. 1)
• Motor rotation direction switch (Fig. 1-1)
• Instrument selector (motor/ultrasound) (Fig. 1-2)
• Power control (Fig. 1-3)
• Spray selector ON/OFF switch (Fig. 1-4)
• Spray control (Fig. 1-5)
• Syringe cord (Fig. 1-6)
• Micro-motor cord (Fig. 1-7)
• Ultrasonic handpiece cord (Fig. 1-8)
b) Rear panel (Fig. 2)
• Connector for air supply (Fig. 2-1).
• Connector for water supply (Fig. 2-2).
• Team Up connection socket (Fig. 2-3).
• Footswitch socket (Fig. 2-4).
• ON/OFF switch (I/O) (Fig. 2-5).
• Power socket (Fig. 2-6).
• Fuse housing (Fig. 2-7).
c) Technical Data
!
Electrical supply: 115/230 VAC ±10% - 50/60 Hz
Class I equipment, type BF
Fuses: 115V: 2 x 1 ATD; 230V: 2 x 0.5 ATT
Frequency of vibrations: 28 to 36 KHz
Water entry pressure: 1 to 5 bar
(14.5 to 72.5 PSI)
Water filter: approx. 60 µm
Motor: (see instructions supplied with the
handpiece):
IS40: 1,600 to 40,000 rpm, internal spray
IMLN: 1,600 to 40,000 rpm, internal spray + light
Overall dimensions:
Height: 260 mm
Width: 280 mm
Depth: 100 mm
Weight: 3.8 Kg
Manufactured in accordance with Directive
93/42/CEE.

1. 3 INSTALLATION AND USE

a) Recommendations
Note: before using for the first time, clean,
decontaminate and sterilise the accessories.
Before using the unit for the first time it is
recommended that you plug in the accessories
and operate the spray for several minutes in order
to rinse the equipment's internal water circuits.
Check that the equipment is correctly plugged in
and isolated; the mains plug with earthing pin
must be used.
The user must check the accessories and cords
regularly for faults in isolation and replace them
if necessary.
Do not use the equipment if it appears to be
faulty.
If the equipment develops a fault, it is
recommended that you contact the supplier of
your Cocoon Hygienist rather than using a non-
specialist repairer.
b) Connections
The electrical connexion must conform to current
norms and the air supply system must meet
quality criteria compatible with dental hygiene.
Rear panel of C
H
OCOON
YGIENIST
Connect :
• the blue flexible tube (Fig. 2-1) to the
compressed air;
• the green flexible tube (Fig. 2-2) to the water
supply;
• the mains cord should be connected to its
socket (Fig. 2-6) and the plug should be earthed
(depending on country).
7
: