The I
C
2 is prepared for use as follows:
MPLANT
ENTER
- Make sure that the device is connected correctly
and properly insulated.
- Switch the power switch (Fig. 2-5) to position I
(ON).
- Suspend the bottles or bags of physiological saline
or sterile water from the brackets.
- Raise the irrigation pump covers.
- Insert the irrigation line cassettes horizontally into
their housings.
- Close the covers.
- Insert the irrigation line perforating pins into the
bottles or pouches of physiological saline and open
the cap on the pins.
- Make any necessary setting adjustments (irrigation
flow rate, speed, torque, power, power mode,
program, etc.) (refer to chapter 7).
Note: This sequence must be completed under
conditions of controlled asepsis for the safety of your
patients. An operating assistant may be needed.
Note: The data is saved automatically, in order to
maintain the display parameters as previously used
upon switchover from the various modes.
I-S
L
function
URGE
ED
- Check that there are no traces of moisture on the I-
S
L
micro-motor or micro-motor cord
URGE
ED
connections. If there are, remove by wiping and
drying with the multifunction syringe fed with
medical-quality filtered air.
- Connect the micro-motor cord to the lefthand
connector of the device (Fig. 1-7).
- Align the micro-motor and cord connector electrical
contacts, then screw the motor onto the connector.
- Connect a contra-angle to the I-S
motor.
- Fit an instrument (drill, bur, etc.) to the contra-
angle.
- Fasten the irrigation line to the cord using line
clips.
- Connect the end of the irrigation line to the contra-
angle.
- Run the motor fitted with its contra-angle and its
instrument without irrigation at moderate speed for
18
10 to 15 seconds in order to diffuse the lubricant
and eliminate excess.
- Prime the irrigation circuit by pressing the control
keypad flush/prime button (Fig. 3-1).
- When the irrigation solution reaches the end of the
contra-angle, release the footswitch.
- Use I
practice.
P
IEZOTOME
- Connect the P
righthand connector of the device (Fig. 1-9).
- Screw the selected tip onto the handpiece using a
tip wrench (according to the recommendations of
the clinical booklet).
- Fasten the irrigation line to the cord using line
clips.
- Connect the end of the irrigation line to the
P
- Prime the irrigation circuit by pressing the
flush/prime button on the screen (Fig. 3-1) or the
multifunction footswitch (Fig. 11-2).
- When the irrigation solution reaches the end of the
tip, release the footswitch.
- Use the I
dental practice.
N
EWTRON
- Check that there are no traces of moisture on the
N
connections. Remove any moisture by wiping and by
drying with the multifunction syringe fed with
medical-quality filtered air.
- Connect the scaler cord to the righthand connector
of the device (Fig. 1-9).
L
micro-
- Align the electrical contacts of the N
URGE
ED
handpiece and the scaler cord, then plug the
handpiece into the cord.
- Screw the selected tip onto the handpiece using a
tip
recommendations).
- Fasten the irrigation line to the cord using line
clips.
- Connect the end of the irrigation line to the scaler
cord.
C
2 in accordance with good dental
MPLANT
ENTER
function
IEZOTOME
2 L
handpiece.
IEZOTOME
ED
C
MPLANT
ENTER
function
L
handpiece or the scaler cord
EWTRON
ED
wrench
(according
2 L
handpiece cord to the
ED
2 in accordance with good
EWTRON
to
the
L
ED
T
IPBOOK