FAULT CODE
FLASH POSSIBLE CAUSE
10 + S
10 + A
Rev. 2.0
The batteries need charging, or there is a bad connection to the battery. Check the
1
connections to the battery. If the connections are good, try charging the batteries.
The left hand motor has a bad connection. Check the connections to the left hand motor.
2
The left hand motor has a short circuit to a battery connection. Contact your Sunrise Medical
authorised dealer.
3
The right hand motor has a bad connection. Check the connections to the right hand motor.
4
The right hand motor has a short circuit to a battery connection. Contact your Sunrise Medical
authorised dealer.
5
The wheelchair is being prevented from driving by an external signal. One possibility is that the
battery charger is plugged in.
6
A joystick fault is indicated. Make sure that the joystick is in the centre position before
switching on the control system.
7
A control system fault is indicated. Make sure the control system connections are secure.
8
The parking brakes have a bad connection. Check the parking brake and the motor
connections. Make sure that the control system connections are secure.
9
An excessive voltage has been applied to the control system. This is usually caused by a poor
battery connection. Check the battery connections
10
S = Speed indicator LED's
A communication fault is indicated. Make sure that the joystick cable is securely
connected and not damaged.
A = Actuator LED's
An actuator trip is indicated. If more than one actuator is fitted, check which actuator
is not working. Check the actuator wiring.
Fig.8.2
31
Jive M2 Sedeo Ergo