10 - TROUBLE-SHOOTING CHART
PROBLEMS
POSSIBLE CAUSES SOLUTIONS
• No air or
• Flow generator off
insufficient air
• Flow generator
comes out of the
regulated wrongly
mask
• Patient circuit tube
not inserted
• The mask is
• Nasal pillows
noisy
mounted incorrectly
and air leak
• The nasal
• You are using the
pillows do not
wrong size
fit into the nasal
cavity
• Exhaled air does
• Vent ports
not flow out and
obstructed
difficulty inhaling
11 - DISPOSAL INSTRUCTIONS
The components of the mask are made of the following materials
and must be disposed of in accordance with the applicable laws.
COMPONENT
Nasal pillows
Main body
Elbow
Tube
Connector
Swivel fitting
Headgear
Tube strap
Rear buckle
Top buckle
• Switch on the flow
generator
• Regulate the flow
generator
• Connect the tube
to the flow generator
and mask
• Mount the cushions
correctly as instructed
• Replace with a more
suitable size (sizes
available S - M - L)
• Remove the
obstruction
MATERIAL
Silicone
Nylon
Polycarbonate
Metallocene
Acetalic
Polycarbonate
Nylon/Polyurethane
Nylon
Nylon
Nylon
ENGLISH
12 - TECHNICAL DATA
MINIMUM CONTINUOUS FLOW
REMARK: Due to manufacturing differences, the exhalation flow
rate may differ from the values shown in the table (±10%).
Normative references
Resistance
A-weighted sound
pressure level
according to ISO 3744
A-weighted sound
power level according
to ISO 3744
Operating pressure
Temperature range
Dead space (mask and
tube assembly)
11
CE
marking
in
accordance
European Directive 93/42/EEC, class
IIa and subsequent amendments. EN
ISO 14971, ISO 5356-1, ISO 10993-1,
EN ISO 17510-2
at 50 lpm: 0,6 cmH
O - at 100 lpm: 2
2
cmH
O
2
11,5 dB(A) ± 1,5 dB(A)
18,2 dB(A) ± 1,5 dB(A)
4 – 40 cmH
O
2
5 – 40 °C
81 ml
with