Table 1 lists the available TS-600 pump models and
indicates the power cord to be used with each particular
pump.
Pump Model
Motor Voltage
PTS06001UNIV
100-115 VAC
PTS06001UNIVEU
200-230 VAC
PTS06001UNIVUK 200-230 VAC
PTS06003UNIV
200-230 VAC
PTS06101UNIV
100-115 VAC
PTS06001UVPI
100-115 VAC
Circuit Breakers
WARNING
Protect against short circuits by
installing a circuit breaker of the
proper capacity.
The circuit breaker must be suitably
located and easily reached.
The circuit breaker must be marked
at the disconnecting device for the
equipment with the on/off symbol
I/O and meet the relevant require-
ments of IEC60947-1 and
IEC60947-3.
Table 2 lists the circuit breaker capacity required for pump
motors of various voltages, single and three-phase.
Table 2

Circuit Breaker Requirements

Single Phase Motor
100-115 VAC
Breaker capacity
20 Amperes
Three Phase Motor
200-230 VAC
Breaker capacity
10 Amperes
TriScroll 600 Series Vacuum Pump
Table 1

Pump Models and Power Cords

Plug Type
NEMA 5-15P
CEE 7/7
BS 1363
-
NEMA 5-15P
NEMA 5-15P
200-230 VAC
50/60 Hz
50/60 Hz
10 Amperes
380-415-460 VAC
5 Amperes
Three phase pumps are shipped without a power cord.
A three phase power cable kit, Agilent part number
S4789001, is available. This kit does not include a plug.
Fusing
Conductor
Size
-
14 Gage
-
3x1.50 mm
BS 1362
3x1.50 mm
13 Amp
-
-
-
14 Gage
-
14 Gage
Cord Marking
UL/CSA
2.5 m
2
<HAR>
2.5 m
2
<HAR>
2.5 m
-
-
UL/CSA
2.5m
UL/CSA
2.5m
Cord
Length
3