Instrument alarm

Intermittent alarm:
Special symbol »
« displayed on the right side
of the display:
– The instrument is not ready for operation.
 Contact maintenance or Draeger Service to rectify the problem.
Activating the Info mode
 In measuring mode, press the [OK] key for approx. 3 seconds.
– If any warning or fault messages exist, the corresponding note or error codes
are displayed (see Technical Handbook).
Press the [OK] key successively for the next display. The peak values and
the exposure values TWA and STEV will be displayed.
– If no key is pressed for 10 seconds, the instrument returns automatically to
measuring mode.
Info Off mode
Press the [ + ] key when the instrument is turned off. The name of the gas,
measuring unit, and measuring range limit value are displayed for all channels.
Press the [ + ] key again to exit the Info Off Mode (or via timeout).

Calling the Quick Menu

 In measuring mode, press the [ + ] key three times.
– If functions in the quick menu are activated using the PC software "Dräger CC
Vision", you can select these functions using the [ + ] key. If no functions have
been activated in the quick menu, the instrument remains in measuring mode.
Possible functions:
1. Bump test (configuration for bump test, see
technical manual)
2. Fresh air calibration
3. Delete peak values
4. Display pump information, see technical manual
5. Activate or deactivate pump, see technical manual
 [OK] key to retrieve the retrieve the selected function.
 Press the [ + ]- key to cancel the active function and to switch to measuring
mode.
– If no key is pressed for 60 seconds, the instrument returns automatically to
measuring mode.
Replacing the batteries / rechargeable batteries
Explosion hazard! To reduce the risk of ignition of a flammable or explosive
atmosphere, strictly adhere to the following warning statements:
Do not throw used batteries into fire or try to open them by force.
Do not replace or charge batteries in potentially explosive areas.
Do not use new batteries with used batteries, and do not mix batteries from
different manufacturers or of different types.
Remove batteries before maintenance work.
Batteries / rechargeable batteries are part of the Ex approval.
Only the following types may be used:
– Alkaline batteries – T3 – (non rechargeable!)
Panasonic LR6 Powerline,
1)
Varta Type 4106
(power one) or
1)
Varta Type 4006
(industrial)
– Alkaline batteries – T4 – (non rechargeable!)
1)
Duracell Procell MN1500
– NiMH rechargeable batteries – T3 – (rechargeable)
1)
GP 180AAHC
(1800) max. 40 °C ambient temperature.
Only charge NiMH power packs T4 (type HBT 0000) or T4 HC (type HBT 0100)
with the appropriate Dräger charger. Charge NiMH single cells for ABT 0100
battery holder as directed by the manufacturer. Ambient temperature during
charging: 0 to +40 °C.
1) Not part of the BVS10 ATEX E 080X and PFG 10 G 001X technical suitability test.
Switching off the instrument:
 Press and hold the [OK] and [ + ] keys simultaneously.
 Loosen the screw on the power pack and remove the power pack.
Battery holder (order no. 83 22 237):
 Replace alkaline batteries or NiMHy rechargeable batteries. Ensure correct
polarity.
NiMH power pack T4 (Type HBT 0000) / T4 HC (Type HBT 0100):
 Replace complete power pack.
 Insert the power pack into the instrument and tighten the screw, the instrument
switches on automatically.