Technical data

Excerpt See the Technical Handbook for details
Ambient conditions:
During operation and
-20 to +50 °C with NiMH power pack types:
storage
HBT 0000 and HBT 0100,
alkaline single cell type:
Duracell Procell MN 1500
-20 to +40 °C for NiMH single cell type:
GP 180AAHC
Panasonic LR6 Powerline
0 to +40 °C for alkali single cell types:
Varta 4006
700 to 1300 hPa
10 to 90 % (to 95 % short-term) rel. hum.
Temperature range over
-40 to +50 °C
2)
a short period
:
Maximum of 15 minutes with NiMH power pack T4
(HBT 0000) or T4 HC (HBT 0100)Requirement:
storage of the instrument at room temperature
(+20 °C) for at least 60 minutes in advance.
Position of use
any
Storage time
X-am 5000
1 year
Sensors
1 year
Electrical classification
IP 67 for instrument with sensors
Alarm volume
Typically 90 dB (A) at 30 cm distance
Operating time:
– Alkaline battery
Typically 12 hours under normal conditions
– NiMH-
power pack:
T4
(HBT 0000)
Typically 12 hours under normal conditions
T4 HC
(HBT 0100)
Typically 13 hours under normal conditions
Dimensions:
approx. 130 x 48 x 44 mm (H x W x D)
Weight:
approx. 220 to 250 g
Refresh interval for
1 s
display and signals:
1)
2)
2
and alkaline single cell type:
2)
2)
, Varta 4106
,
CE marking:
See declaration of conformity on page 328.
3)
Approvals
:
(see "Notes on Approval" on page 327)
The BVS 10 ATEX E 080 X technical suitability test
is based on the adjustment with the target gas.
1) The Technical Handbook and Instructions for Use / data sheets for the utilized sensors and the PC
software CC Vision for Dräger X-am 5000 can be downloaded on the product page of the X-am
5000 at the following Internet site: www.draeger.com See also the instructions for use provided and
the data sheets for the sensors used.
2) Not part of the BVS10 ATEX E 080X and PFG 10 G 001X technical suitability test.
3) The technical suitability tests are valid for the X-am 5000 gas detection instrument and the
calibration cradle. The explosion-protection approvals are only valid for the X-am 5000 gas
detection instrument; the calibration cradle must not be used in the Ex zone.