Use
For mixing and agitating liquids.
IKA
Roller 6/10 basic
®
- Continuous mode for single or multiple test tubes.
IKA
Roller 6/10 digital
®
- Continuous mode for single or multiple test tubes.
- Continuous mode for single or multiple test tubes.
Individual rollers can be removed to allow tubes of different sizes
to be processed. In this case the rollers automatically take up
symmetrical positions on the shaking table, so that the rollers
can be removed and inserted as desired.
plate.
supply the device with power. See the section "Technical data"
for the connection values.
IKA
Roller 6/10 basic
®
The AC adapter is suitable for 100 V upto 240 V. Insert an appropriate
adapter corresponding to the country, before plugging the AC adapter
into the mains socket.
Switching the device on
Press the "Power" button (L) to switch the device on.
The LED (M) lights up green.
Starting the device
Press the "Start/Stop" button (F) to start the shaking process.
The device starts at the preset speed of 30 rpm.
Stopping the device
Press the "Start/Stop" button (F) to stop the shaking process.
Switching off the device
Press the "Power" button (L) to switch the device off.
The LED (M) goes out.
Correct use
- Laboratories
- Pharmacies
The unit is suitable for use in all areas apart from domestic areas
and areas directly connected to a low-voltage supply which also
serves domestic areas.
The safety of the user cannot be guaranteed if the appliance is
operated with accessories that are not supplied or recommended
by the manufacturer or if the appliance is operated improperly
Useful information
The integral tube stopper prevents the tubes falling out.
For the safety of the operator, the device switches off automati-
cally if the rollers jam and/or can no longer rotate freely.
Commissioning
IKA
when power is restored after a power outage. If the IKA
Roller 6/10 digital is running in mode A it does not restart
automatically when power is restored after a power outage.
- Schools
- Universities
Roller 6/10 basic does not restart automatically
®
If these conditions are met the unit is
ready for service when the mains plug
has been plugged in.
®
12