The AC adapter is suitable for 100 V upto 240 V. Insert an appropriate
adapter corresponding to the country, before plugging the AC adapter
into the mains socket.
Switching the device on
Press the "Power" button (L) to switch the device on.
The following data appear on the display in the sequence listed:
- the test image
- the software version
- the operating mode
- the values and settings last used
Speed setting
Use the "Speed setting" button (K) to set the speed to a value between 5
and 80 rpm in units of 1 rpm.
Timer/Counter
Select "Counter" or "Timer" (I).
Timer
When "Timer" is selected, the set value counts down. Once the timed
shaking cycle has been completed the device switches off automatically.
An acoustic signal beeps three times and the "Timer" LED (H) fl ashes red.
Counter
The counter measures the time from the start of shaking to the end.
The "Counter" LED (G) fl ashes red.
If the "Start/Stop" button (F) is pressed the shaking process stops and
the elapsed time is displayed. If the counter is restarted the measurement
restarts afresh.
To switch from counter to timer or vice versa, press the "Timer" button (I).
Setting
Set the timer from 1 sec to 99 h 59 min using the "Plus/Minus" timer
setting buttons (J).
Press the "Start/Stop" button (F) to start the shaker.
The last timer value setting is saved and is loaded for the next shaking
process.
Resetting to 0
To reset the timer, press the "Timer Minus" and "Timer Plus" buttons (J)
simultaneously.
Starting the device
Press the "Start/Stop" button (F) to start the shaking process.
The device starts with the desired settings.
Stopping the device
Press the "Start/Stop" button (F) to pause operation.
Switching off the device
Press the "Power" button (L) to switch the device off.