Trip procedure

When a dysfunction occurs on the screw vacuum pump, it should
immediately be stopped. Close without delay the suction valve MV1, in
order to avoid the ventilation of the recipient via the screw vacuum
pump and the pump has to be immediately inertised.

Shutdown procedure

l A problem has been indicated by a monitoring element during the
running of the pump
l Close immediately the suction valve MV1 and simultaneously
supply non explosive gas into the flushing gas connection, suction
side
l Switch off the vacuum pump (motor)
l Non explosive gas volume flow:
u NC 0100 B : ³ 25 liter/ min
u NC 0200 B: ³ 40 liter/ min
u NC 0300 B : ³ 75 liter/ min
l Inertise during at least 200 seconds, then close the gas supply and
the inerting is finished
l Switch off the pump, close the solenoid valves MV2, MV3, MV5,
MV7
Before to restart the vacuum pump, the cause of the problem has to be
found and solved. This is particularly important when TSA+/ 0102,
TSA+/ 0103, PSA+/ 0301 or SA+/ 0701 have indicated the problem.
The vacuum pump should be installed such a way than when the
pump stops, a rear current resp. a reflux of the drawn medium is not
possible.

Protection against explosions

The vacuum pumps COBRA NC 0100 ¼ 0300 B are conformable to
the European directive 2014/ 34/ EU.
The vacuum pumps are provided with one of the following
characteristics for the protection against the explosions:
COBRA NC 0100 B
Inside
II
2G
2G
Outside
II
1
2
3
4
1
Apparatus according to directive 2014/ 34/ EU
Inside and outside of the vacuum pump
2
Equipment group II (Not for mining application)
3
Possible explosive gas atmosphere in normal running inside and outside
(apparatus category 2, for Zone 1)
4
5
Type of protection
Explosion group
6
7
Maximum gas and surface temperature (temperature class T4 = 135 °C, T3 = 200 °C)
Equipment protection level
8
Special conditions for safety utilization
9
Flushing gas
connection
Gb
X
Ex h
IIC
T4
Gb
X
Ex h
IIC
T4
8
6
7
9
5
For the adjustment of the temperature class, imperatively refer to
"Adjustment of the thermostatic valve (option) and adjustment of the
temperature class".
In the table below, you will find the possible alternative of temperature
classes depending of the cooling system and the maximal motor
frequency:
Cooling
Direct cooling
system
Temperature
Max. motor
Type
class
frequency
NC 0100 B
T4 (i/o)
60 Hz
COBRA NC 0200 B
Inside
II
2G
2G
Outside
II
The following temperature classes are available
T4 (i/o)
or
T3 (i)/ T4(o)
or
T3 (i/o)
when the vacuum pump is equipped with a direct cooling system. The
internal and external temperature class must be identical! If the
vacuum pump is equipped with a recirculating cooling system, only the
temperature class:
T3 (i)/ T4(o)
or
T3 (i/o)
is available.
When the vacuum pump is equipped with a direct cooling system,
imperatively refer to "Adjustment of the thermostatic valve (option)
and adjustment of the temperature class".
In the table below, you will find the possible alternative of temperature
classes depending of the cooling system and the maximal motor
frequency:
Cooling
Direct cooling
system
Temperature
Type
class
T4 (i/o)
T3 (i)/ T4(o)
NC 0200 B
T3 (i/o)
COBRA NC 0300 B
II
2G
Inside
2G
II
Outside
The following temperature classes are available
T4 (i/o)
or
T3 (i/o)
when the vacuum pump is equipped with a direct cooling system. The
internal and external temperature class must be identical! If the
vacuum pump is equipped with a recirculating cooling system, only the
temperature class T3 (i/o) is available.
Recirculating cooling
Temperature
Max. motor
class
frequency
T4 (i/o)
50 Hz
T3, T4*
Gb
Ex h
IIC
Ex h
T3, T4*
IIC
Gb
Recirculating cooling
Max. mo-
Max. mo-
Tempera-
tor fre-
tor fre-
ture class
quency
quency
50 Hz
T3 (i/o)
60 Hz
60 Hz
T3 (i)/
60 Hz
T4(o)
60 Hz
T3, T4*
IIC
Gb
Ex h
Ex h
T3, T4*
IIC
Gb
Description
Seite/ Page/ Page 13
X
X
X
X

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Cobra nc 0200 bCobra nc 0300 b

Table des Matières