13.4 Error messages

Error message
ERR4
ERR5
ERR6
Should other error messages occur, switch balance off and then on again. If the error
message remains inform manufacturer.
GAB-N-BA-e-1817
Description
Zeroing range exceeded
due to switching-on balance
0
or pressing
(normally
4% max)
Keyboard error
Value outside the A/D
changer range
Unable to initialise zero
point
Possible causes
• Object on the weighing
plate
• Overload when zeroing
• Improper adjustment
• Damaged weighing cell
• Damaged electronics
• Improper operation of the
balance
• Damaged weighing cell
• Damaged electronics
• Measuring cell defective
/ overloaded
• Object on weighing pan
/ contact
• Transport safety device
has not been removed
• Main board defective
39