Explanation Of The Data Transfer; Remote Control Instructions - KERN and SOHN GAB-N Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

12.1.3 Explanation of the data transfer

Example:
Header1
,
Header2
Header 1
2 byte, St or US, ST=weighing value stable, US=weighing value
instable
Header 2
2 byte, G or N, G=gross weight, N= net weight
space
Blank
W1-W7
Weight value with decimal position
Unit
2 byte, kg or lb
Terminator
<CR> <LF> (CR=Carriage return)

12.2 Remote control instructions

The remote control commands are sent from the remote control unit to the balance
as ASCII code. After the balance having received the s/w/t commands, it will send the
following data.
Take into account that the following remote control commands must be sent without a
subsequent CR LF.
T: tare
Taring
Z: zero
Zeroing
S: stable status
Sending stable weight values
W:
Sending stable or unstable weight values
P:
Sending the number of pieces (PCS)
GAB-N-BA-e-1817
,
-/space
W1
W2
(LF= Line Feed)
W3
W4
W5
W6
W7
,
unit
terminator
37

Publicité

Table des Matières
loading

Table des Matières