The Mistral Nebulizer aerosol does not need maintenance or lubrication.
With regard to training, given the information contained in the user manual and since it is easy to understand the said
device, it doesn't appear to be necessary. It is necessary to check functioning and instrument before every use. Unpack
the instrument and always check integrity of plastic parts and feeding cable., they might have been damaged during
previous use.
Connect cable to electrical network and turn switch on. Close the compressor mouthpiece with a finger and verify that
loud noises are not present, these can indicate wrong functioning. Verify that the atomiser is not damaged by previous
use (it was badly put away or badly knocked).
A protection fuse (F 2A L 250V) not reachable from exterior protects the instrument.
For fuse replacing, please make reference to manufacturer technical personnel.
GIMA will provide upon request electric diagrams, components list, description, setting instructions and any other
information that can help the technical assistance staff for product repair.
Fault type
1. Low Nebulization
2. Low Nebulization
3. Absence of Nebulization
4. Slow Nebulization
5. The device doesn't work
6. Noisy Device
Fault 1 - 2 - 3 - 4 - 5 - 6
service GIMA.
If the unit doesn't nebulizer once the above conditions have been checked, we suggest to contact your dealer or technical
BEFORE EVERY CHECKING OPERATION, IN CASE OF ANOMALIES OR BAD FUNCTIONING, PLEASE CONTACT GIMA
TECHNICAL SERVICE. GIMA DOES NOT GIVE GUARANTEE IF INSTRUMENT, AFTER THE TECHNICL SERVICE
ANY MINIMAL MODIFICATION/REPAIR ON THE DEVICE VOIDS THE WARRANTY AND DOES NOT GUARANTEE THE
COMPLIANCE WITH THE TECHNICAL REQUIREMENTS PROVIDED BY THE MDD 93/42/EEC DIRECTIVE (AND
Before using and/or after cleaning, pay special attention to ensure that all the accessories supplied with the device are
intact. Switch off the device before cleaning it and disconnect the mains cable from the electrical socket.

PREPARATION

1.
Pull out the air tube from the nebulizer and leave it plugged into the air outlet nozzle of the device;
2.
Rotate the upper part of the nebulizer anti-clockwise;
3.
Use your fingers to disconnect the internal pisper at the bottom of the nebuliser.
CLEANING
Before and after each use proceed with cleaning all of the components of the nebulizer (with the exception of the air
tube) according to one of the two methods described below.
Method 1: Thoroughly clean the components for 5 minutes, using warm drinking tap water (about 40°C) and/or mild
soap.
Method 2: Clean the components (except for the air tube) by immersing them in a solution with 60% water and 40%
white vinegar. When finished, thoroughly rinse with warm drinking water (approx. 40°C).
After cleaning, rinse thoroughly by removing the excess water and allow to air dry in a clean place.
MAINTENANCE
Cause
Clogged Nebulizer Tank
Clogged Nebulizer Tank
Clogged Nebulizer Tank
Highly dense drug
Power Cable is damaged
External power source failure
Extended use
No solution with previous items
CHECKING, APPEARS TO BE TAMPERED.
SUBSEQUENT CHANGES) AND ITS NORMATIVES
CLEANING AND WASHING OF ACCESSORIES
Clean and disinfect the nebulizer tank as
explained in the instruction manual
If cleaning was not succesful change cruet
Check that the nebulizer contains medication;
Make sure that the nebulizer is not clogged;
Check the connection between the
compressor air outlet port and the
Dilute drug in physiological liquid
Replace the power cable
Call retainer or manufacturer GIMA
Check the external power source
Call retainer or manufacturer GIMA