Eclipse 5
Alarms Table
Alarm
All OK, warm-up done
Warm-up O2 < 85%
No breath detected (less than
45 seconds)
No breath detected (exceeding
45 seconds)
One wire communication loss
Cannot charge battery
Ambient pressure fail
O2 < 85% (After warm-up)
Flow rate error > +/-10%
Warm battery > 60°C
Low battery
(5% remaining battery
capacity)
Loss of power
Malfunction
20 - ENG
PN 21182453-C4 D | User Manual
LCD Text
Green
Yellow
NO ALARM
ON
OFF
O2 LOW
ON
ON
NO ALARM
ON
OFF
P <--> C
ON
ON
ONE WIRE
ON OR
ON
OFF
CHARGER
ON OR
ON
OFF
AMB PRESS
ON OR
ON
OFF
O2 LOW
OFF
ON
FLOWRATE
OFF
ON
WARM BAT
ON
ON
LOW BAT
ON
ON
********
OFF
ON
FAIL XX
OFF
ON
Audio
Code
OFF
000
No Beeps
008
OFF
000
1 alarm beep, repeat
200
every 20 seconds
1 alarm beep, repeat
100
every 20 seconds
1 alarm beep, repeat
001
every 20 seconds
1 alarm beep, repeat
004
every 20 seconds
1 alarm beep, repeat
008
every 20 seconds
1 alarm beep, repeat
020
every 20 seconds
1 alarm beep, repeat
002
every 20 seconds
1 alarm beep, repeat
010
every 20 seconds
1 alarm beep, repeat
040
every 20 seconds for
200 seconds
1 alarm beep, repeat
080
every 20 seconds for
200 seconds