Eclipse 5
Pulse Dose Mode and autoSat Features
Your Eclipse 5 has a feature, called autoSAT
delivers a consistent-sized bolus of oxygen up to
40 breaths per minute (see Max Breath Rate table
below). As your breath rate increases, ie. during
ambulation, the autoSAT feature servo-controls
the device, delivering the set bolus size. autoSAT
Technology is proprietary to CAIRE and is utilized
during Pulse Mode only.
Max Breath Rate
AC Power
Bolus
Supply &
Size
Pulse
Power
Dose
Cartridge
Setting
(Battery)
15%)
Max Breath
mL
Rate
1.0
16
40
2.0
32
40
3.0
48
40
4.0
64
40
5.0
80
37
6.0
96
31
7
128
23
8
160
18
9
192
15
Note: Bolus volume decreases as breath rate
exceeds published range.
Note: The pulse dose setting number (1-9) is not
equal to LPM.
WARNING: PULSE DOSE MODE
SETTINGS MUST BE DETERMINED
FOR EACH PATIENT INDIVIDUALLY
FOR THEIR NEEDS AT REST, DURING
EXERCISE AND WHEN TRAVELING.
WARNING: THE ECLIPSE 5 MAY NOT BE ABLE TO
DETECT ALL INSPIRATORY EFFORTS IN PULSE
MODE. IF THE ECLIPSE 5 DOES NOT RESPOND TO
YOUR INSPIRATION EFFORTS, ASK YOUR HOME
CARE PROVIDER TO CHECK THE SENSITIVITY
SETTING.
18 - ENG
PN 21182453-C4 D | User Manual
Note: To Equipment Provider: The following oxygen
that
®
administration accessories are recommended for
use with the Oxygen Concentrator:
• Nasal Cannula with 7 feet (2.1 m) of tubing: PN
5408-SEQ
• For any additional recommended accessories,
please see the Accessories Catalog (PN ML-
LOX0010) available on www.CAIREmedical.com.
A firebreak is required for use with any cannula.
• CAIRE offers the OxySafe™. This is intended to be
used in conjunction with the oxygen concentrator.
DC Power
The OxySafe™ is a thermal fuse to stop the flow of
Supply
gas in the event that the downstream cannula or
oxygen tubing is ignited and burns to the OxySafe.
Max Breath
It is placed in-line with the nasal cannula or oxygen
Rate
tubing between the patient and the oxygen outlet
40
of the Eclipse 5.
40
For proper use of the OxySafe, always refer to the
40
manufacturer's instructions (included with each
31
OxySafe kit).
25
20
CAIRE offers an OxySafe™ kit that includes a
15
cannula headpiece, OxySafe™, and 7' tubing: PN
20629671.
12
10
Step 5:
the Prescribed Level
setting to the prescribed setting supplied by your
physician.
WARNING: IT IS VERY IMPORTANT
TO SELECT ONLY THE PRESCRIBED
LEVEL OF OXYGEN. DO NOT CHANGE
THE FLOW SELECTION UNLESS YOU
HAVE BEEN DIRECTED TO DO SO BY A
LICENSED CLINICIAN.
Step 6: Begin Using Your Eclipse 5
Breathe normally through the nasal cannula.
Step 7: Universal Cart Operation
The Universal Cart is designed for use on pedestri-
an-type surfaces.
Align the locater holes on the bottom of the device
to the tabs on the Universal Cart.
Align the large threaded screw
to the insert at the back of the
Eclipse 5. Then hand-tighten
the knob on the cart until the
Universal Cart is tightly secured
Adjust the Flow Setting to