to the device.
Depress the push-button on the
Universal Cart handle to adjust
the height of the handle.
Step 8:
Power Off the Device
Press and hold the "ON/OFF" Button for two (2)
seconds to power OFF your Eclipse 5.
User Troubleshooting Table
Do Not Ignore Alarms
WARNING: THE MANUFACTURER RECOMMENDS AN ALTERNATE SOURCE OF SUPPLEMEN-
TAL OXYGEN IN THE EVENT OF A POWER OUTAGE, ALARM CONDITION, OR MECHANICAL
FAILURE. CONSULT YOUR PHYSICIAN OR EQUIPMENT PROVIDER FOR THE TYPE OF RESERVE
SYSTEM REQUIRED.
Symptom
Possible Cause
The Eclipse 5 has de-
The yellow light is
tected a problem with
on solid. There is no
the connection to the
audible alarm.
power cartridge.
The yellow light is
The Eclipse 5 power
on. Battery symbol is
cartridge may be
flashing on LCD.
warm.
The yellow light is
The Eclipse 5 power
on. Battery symbol is
cartridge voltage may
flashing on LCD.
be low.
The Eclipse 5 may
The yellow light is on.
have detected a flow
The green light is off.
restriction causing
low or blocked flow.
Store the Eclipse 5 in a cool and dry location for a
period no longer than 1 month. The Eclipse 5 must
run for a minimum of 2 hours per month to ensure
proper function of the unit.
Note: DO NOT operate the Eclipse 5 and ambulate
while a humidifier is attached. Remove the humidi-
fier bottle before walking. DO NOT lay the Eclipse 5
down while attached to a humidifier bottle.
Yellow Light
Your Action
1) Remove power cartridge and re-install to ensure that it is
secure in the concentrator.
2) If possible, install a separate power cartridge in the Eclipse
5.
3) If this problem persists, contact healthcare provider.
1) Re-attach the Eclipse 5 to external power or install a
fully-charged power cartridge in the Eclipse 5. Allow warm
power cartridge to cool outside the concentrator for 30
minutes.
2) Re-charge the power cartridge using AC power
3) Replace the power cartridge if another is available.
4) If the condition persists, contact your healthcare provider.
1) Plug Eclipse 5 into AC power supply and recharge or replace
power cartridge with fully charged power cartridge.
1) Clean and replace cabinet inlet filter
2) Ensure that the cannula is not kinked or blocked. If used
with a humidifier bottle, ensure that it is filled properly and
not creating a blockage.
3) Ensure that the Eclipse 5 has proper ventilation. It needs to
be at least 3 inches from any surface to ensure the vents are
not blocked.
4) If the problem persists, switch to an alternate source of
oxygen and contact healthcare provider for assistance.
PN 21182453-C4 D | User Manual
Eclipse 5
19 - ENG